Introduction: Computed tomography (CT) is the preferred diagnostic toolkit for head and brain imaging of head injury. A recent development is the invention of a portable CT scanner that can be beneficial from a clinical point of view. Aim: To compare the quality of CT brain images produced by a fixed CT scanner and a portable CT scanner (CereTom). Methods: This work was a single-centre retrospective study of CT brain images from 112 neurosurgical patients. Hounsfield units (HUs) of the images from CereTom were measured for air, water and bone. Three assessors independently evaluated the images from the fixed CT scanner and CereTom. Streak artefacts, visualisation of lesions and grey–white matter differentiation were evaluated at three different levels (centrum semiovale, basal ganglia and middle cerebellar peduncles). Each evaluation was scored 1 (poor), 2 (average) or 3 (good) and summed up to form an ordinal reading of 3 to 9. Results: HUs for air, water and bone from CereTom were within the recommended value by the American College of Radiology (ACR). Streak artefact evaluation scores for the fixed CT scanner was 8.54 versus 7.46 (Z = -5.67) for CereTom at the centrum semiovale, 8.38 (SD = 1.12) versus 7.32 (SD = 1.63) at the basal ganglia and 8.21 (SD = 1.30) versus 6.97 (SD = 2.77) at the middle cerebellar peduncles. Grey–white matter differentiation showed scores of 8.27 (SD = 1.04) versus 7.21 (SD = 1.41) at the centrum semiovale, 8.26 (SD = 1.07) versus 7.00 (SD = 1.47) at the basal ganglia and 8.38 (SD = 1.11) versus 6.74 (SD = 1.55) at the middle cerebellar peduncles. Visualisation of lesions showed scores of 8.86 versus 8.21 (Z = -4.24) at the centrum semiovale, 8.93 versus 8.18 (Z = -5.32) at the basal ganglia and 8.79 versus 8.06 (Z = -4.93) at the middle cerebellar peduncles. All results were significant with P-value < 0.01. Conclusions: Results of the study showed a significant difference in image quality produced by the fixed CT scanner and CereTom, with the latter being more inferior than the former. However, HUs of the images produced by CereTom do fulfil the recommendation of the ACR.
机译:简介:计算机断层扫描(CT)是用于头部和头部颅脑成像的首选诊断工具包。从临床的观点来看,便携式CT扫描仪的发明是一项新的发展。目的:比较固定CT扫描仪和便携式CT扫描仪(CereTom)生成的CT脑图像的质量。方法:这项工作是对112位神经外科患者的CT脑图像进行的单中心回顾性研究。测量了CereTom图像的Hounsfield单位(HUs)中的空气,水和骨骼。三名评估员独立评估了来自固定CT扫描仪和CereTom的图像。在三个不同的水平(中枢半卵,基底神经节和小脑中部花梗)上评估了条纹假象,病变的可视化和灰白色物质的分化。每次评估的得分为1(差),2(平均)或3(好),总和形成3到9的顺序读数。结果:CereTom的空气,水和骨骼的HUs均在美国人的推荐值范围内放射学院(ACR)。固定CT扫描仪的条纹伪影评估得分是在中心半卵形的CereTom为8.54对7.46(Z = -5.67),在基底神经节为8.38(SD = 1.12)对7.32(SD = 1.63)和8.21(SD = 1.30)。对比小脑中轴的6.97(SD = 2.77)。灰白质分化显示,在半卵状卵圆形的得分为8.27(SD = 1.04)对7.21(SD = 1.41),在基底神经节为8.26(SD = 1.07)对7.00(SD = 1.47)和8.38(SD = 1.11)小脑中部花梗则为6.74(SD = 1.55)。病变的可视化显示,在中央半卵状体的分数为8.86对8.21(Z = -4.24),在基底神经节为8.93对8.18(Z = -5.32),在中小脑梗处的分数为8.79对8.06(Z = -4.93)。所有结果均显着,P值<0.01。结论:研究结果表明,固定式CT扫描仪和CereTom产生的图像质量有显着差异,后者比前者差。但是,CereTom生成的图像的HU确实符合ACR的建议。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号