Users of cochlear implants (auditory aids, which stimulate the auditory nerve electrically at the inner ear) often suffer from poor speech understanding in noise. We evaluate a small (intermicrophone distance 7 mm) and computationally inexpensive adaptive noise reduction system suitable for behind-the-ear cochlear implant speech processors. The system is evaluated in simulated and real, anechoic and reverberant environments. Results from simulations show improvements of 3.4 to 9.3 dB in signal to noise ratio for rooms with realistic reverberation and more than 18 dB under anechoic conditions. Speech understanding in noise is measured in 6 adult cochlear implant users in a reverberant room, showing average improvements of 7.9–9.6 dB, when compared to a single omnidirectional microphone or 1.3–5.6 dB, when compared to a simple directional two-microphone device. Subjective evaluation in a cafeteria at lunchtime shows a preference of the cochlear implant users for the evaluated device in terms of speech understanding and sound quality.
机译:人工耳蜗(听觉辅助器,可通过电刺激内耳的听觉神经)的使用者经常会受到噪声中语音理解能力的困扰。我们评估了一个小巧的麦克风之间的距离(7毫米),并且计算价格低廉,适合于耳后人工耳蜗语音处理器的自适应降噪系统。该系统在模拟和真实,无回声和混响环境中进行评估。模拟结果表明,对于具有真实混响的房间,在无声条件下的信噪比提高了3.4至9.3 dB,并且超过了18 dB。在一个混响室内对6位成年人工耳蜗植入用户进行了语音语音理解测试,与单个全向麦克风相比,声音平均理解为7.9–9.6µdB,而与简单定向两麦克风设备相比,则为1.3–5.6µdB。午餐时间在自助餐厅中进行主观评估表明,在语音理解和声音质量方面,人工耳蜗植入用户偏爱被评估的设备。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号