首页> 外文期刊>International Journal of Scientific & Technology Research >Fabrication Of Gold Nanoparticles-Modified Glassy Carbon Electrode And Its Application For Voltammetric Detection Of Cr(III)
【24h】

Fabrication Of Gold Nanoparticles-Modified Glassy Carbon Electrode And Its Application For Voltammetric Detection Of Cr(III)

机译:金纳米粒子修饰玻碳电极的制备及其在伏安法测定Cr(III)中的应用

获取原文

摘要

Abstract: A sensitive and selective differential pulse stripping voltammetric (DPSV) method is presented for the determination of trace amount Cr(III) using glassy carbon electrode (GCE) modified with gold nanoparticles (AuNPs). The method includes AuNPs fabrication and self-assembly GCE modification. This processes replaced the 'H groups with 'NH2 groups on the surfaces of GCE which increased the number of AuNPs attached to it. The GCE modified with AuNPS was used as electrochemical sensor of Cr(III) for the first time that gave a wide linear range from 0.5 to 75 ppb and a very low detection limit of 10 ppt (equivalent to 0.19 nM). The electrode exhibited high reproducibility in repetitive measurements with a relative standard deviation better than 2.4%. The effect of interfering ions study showed that Cu(II), Cd(II), Zn(II) Cr(VI), Ni(II) and Fe(III) ions did not have a significant effect on the measurement.
机译:摘要:提出了一种灵敏且选择性的差分脉冲溶出伏安法(DPSV),用于修饰金纳米颗粒(AuNPs)的玻碳电极(GCE)测定痕量Cr(III)。该方法包括AuNPs制造和自组装GCE修改。该过程在GCE表面上用'NH 2基团代替了'H基团,这增加了附着在其上的AuNP的数量。用AuNPS修饰的GCE首次用作Cr(III)的电化学传感器,其线性范围从0.5到75 ppb,检测限很低,为10 ppt(相当于0.19 nM)。该电极在重复测量中显示出高重现性,相对标准偏差优于2.4%。干扰离子的影响研究表明,Cu(II),Cd(II),Zn(II)Cr(VI),Ni(II)和Fe(III)离子对测量没有显着影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号