首页> 外文期刊>International Journal of Integrated Engineering >Cuff-Less Continuous Blood Pressure Monitoring System Using Pulse Transit Time Techniques
【24h】

Cuff-Less Continuous Blood Pressure Monitoring System Using Pulse Transit Time Techniques

机译:使用脉搏传输时间技术的无袖连续血压监测系统

获取原文
       

摘要

This paper describes the development of a continuous cuff-less blood pressure system based on the pulse transit time (PTT) technique. In this study, PTT is defined by two different approaches denoted as PTT1 and PTT2. PTT1 is the time difference between the R-wave peak of the Electrocardiogram (ECG) and the peak of the Photoplethysmogram (PPG). PTT2 is the time difference between two peak PPG signals on same cardiac cycle at different positions on the body. The ECG is acquired on the chest using 3 lead electrodes and a reflection mode optical sensor is deployed on brachial artery and fingertip to monitor the PPGs. These data were synchronized using a National Instruments data acquisition card along with Matlab software for subsequent analysis. A wrist-type cuff-based blood pressure device was used to measure blood pressure on the right hand. Brachial blood pressure was measured on the upper left arm using oscillometric blood pressure monitor. Experiments were conducted by elevating the right hand at different position to investigate variability of PTT under the effects of hydrostatic pressure. Next the variability of PTT due to blood pressure changes during a Valsalva maneuver was investigated. The result shows that the PTT1 is inversely proportional to blood pressure in both experiments. Meanwhile, there is weak correlation between PTT2 and blood pressure measurement which suggests that by excluding the pre-ejection period (PEP) time in PTT calculation may reduce the accuracy of PTT for blood pressure measurement. In conclusion, PTT measurement between ECG and PPG signals has potential to be a reliable technique for cuff-less blood pressure measurement.
机译:本文介绍了一种基于脉搏穿越时间(PTT)技术的连续无袖血压系统的开发。在这项研究中,PTT由两种不同的方法定义为PTT1和PTT2。 PTT1是心电图(ECG)的R波峰值和光电容积描记图(PPG)峰值之间的时间差。 PTT2是相同的心动周期上人体不同位置的两个峰值PPG信号之间的时间差。使用3个导联电极在胸部获取ECG,并在肱动脉和指尖部署反射模式光学传感器以监测PPG。使用National Instruments数据采集卡和Matlab软件对这些数据进行同步,以进行后续分析。使用腕式袖带血压仪测量右手血压。使用示波血压计在左上臂测量肱动脉血压。通过在不同位置举起右手进行实验,以研究静水压力作用下PTT的可变性。接下来,研究了在瓦尔萨尔瓦(Valsalva)演习中由于血压变化引起的PTT变异性。结果表明,在两个实验中,PTT1与血压成反比。同时,PTT2与血压测量之间的相关性较弱,这表明通过在PTT计算中排除预喷射期(PEP)时间可能会降低PTT用于血压测量的准确性。总之,ECG和PPG信号之间的PTT测量有可能成为无袖血压测量的可靠技术。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号