Analysis of Detector-Source Distance and Detector Bias Voltage for Time Resolution of Plastic Scintillation Detectors

机译:用于塑料闪烁探测器时间分辨的探测器源距离和探测器偏置电压分析

获取原文
           

摘要

The effects of the source-detector distance, the detector bias voltage variation and collimator usage on the time spectra of ~(133)Ba and ~(207)Bi calibration sources were investigated through a spectrometer which was consisted of a BC400 type general purpose plastic scintillation detector. Constant fraction timing method was used to test these effects on the time resolution values. Optimum source-detector distance and bias voltage values with and without collimator were determined for the optimum time resolution. It was concluded that the source to detector apart, the applied bias voltage to the used detector and the collimator usage were quiet effective parameters in coincidence timing experiments.
机译:通过由BC400型通用塑料组成的光谱仪研究了源-检测器距离,检测器偏置电压变化和准直仪使用情况对〜(133)Ba和〜(207)Bi校准源的时间谱的影响。闪烁探测器。使用恒定分数计时方法来测试这些对时间分辨率值的影响。为获得最佳时间分辨率,确定了使用和不使用准直仪的最佳源检测器距离和偏置电压值。结论是,在重合定时实验中,分开检测器的源,向使用的检测器施加的偏置电压和准直器的使用是安静的有效参数。