首页> 外文期刊>Insights into Imaging >Plain radiography in patients treated with intrathecal drug delivery using an implantable pump device
【24h】

Plain radiography in patients treated with intrathecal drug delivery using an implantable pump device

机译:使用植入式泵装置对鞘内给药的患者进行平片检查

获取原文
           

摘要

ObjectivesIntrathecal drug administration using an implanted pump system is well established in intractable spasticity and pain. However, despite continuous advancements in manufacturing technology, adverse events related to the pump and catheter still occur. Most of them, such as migration, damage, disconnection and occlusion, are related to the spinal catheter. The aim of this overview is to update radiologists on how plain radiography of the implanted delivery system for intrathecal drug administration should be interpreted and to increase awareness for the need of urgent and timely multidisciplinary troubleshooting.MethodsPlain radiographic images of patients treated with intrathecal drug administration using an implantable drug delivery system were analysed in a multidisciplinary setting at our (university) referral centre for complications in intrathecal drug administration.ResultsExamples of catheter-related adverse events are described and a proposal is made for stepwise interpretation of standard plain radiographic images.ConclusionsPlain radiological images are the mainstay for the diagnosis of catheter-related adverse events in intrathecal drug delivery. Radiologists play an important role in an early diagnosis. An awareness of abnormal radiological findings seems important to avoid a life-threatening withdrawal syndrome.Teaching points? Untimely cessation of intrathecal drug delivery can lead to a life-threatening withdrawal syndrome.? Initially mild symptoms can lead to an exacerbation of a withdrawal syndrome.? Most intrathecal catheter-related problems are visible on plain radiography.? Common causes of catheter problems are migration, lacerations, occlusion and disconnection.? Knowledge on implanted intrathecal catheters is crucial for interpretation of plain radiography.
机译:目的使用植入式泵系统进行鞘内给药在顽固性痉挛和疼痛方面是公认的。但是,尽管制造技术不断进步,与泵和导管有关的不良事件仍然发生。它们中的大多数(例如迁移,损伤,断开和阻塞)与脊椎导管有关。本概述的目的是为放射科医生提供最新信息,以了解应如何解释鞘内给药的植入式输送系统的普通放射线照相,并提高对紧急和及时的多学科故障排除需求的认识。在我们的(大学)转诊中心的多学科环境中分析了植入式给药系统,以了解鞘内给药的并发症情况。结果描述了导管相关不良事件的示例,并提出了逐步解释标准平片影像学的建议。图像是鞘内给药中导管相关不良事件诊断的主要手段。放射科医生在早期诊断中起着重要作用。认识放射线异常发现似乎很重要,可以避免危及生命的戒断综合征。鞘内给药的不及时停止会导致危及生命的戒断综合征。最初的症状较轻会导致戒断综合征加重。大多数鞘内导管相关的问题在X线平片上可见。导管问题的常见原因是迁移,割伤,阻塞和断开。有关植入式鞘内导管的知识对于解释平片检查至关重要。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号