首页> 外文期刊>Indian Journal of Ophthalmology >Comparison of intraocular pressure measurements using Goldmann tonometer, I-care pro, Tonopen XL, and Schiotz tonometer in patients after Descemet stripping endothelial keratoplasty
【24h】

Comparison of intraocular pressure measurements using Goldmann tonometer, I-care pro, Tonopen XL, and Schiotz tonometer in patients after Descemet stripping endothelial keratoplasty

机译:Descemet剥离内皮角膜移植术后使用Goldmann眼压计,I-care pro,Tonopen XL和Schiotz眼压计测量眼内压的比较

获取原文

摘要

Purpose: Post-Descemet stripping endothelial keratoplasty (DSEK) patients are prone for intraocular pressure (IOP) elevations and glaucoma. Corneal characteristics influence various IOP measuring devices in various ways. The aim of this study was to evaluate the agreement between four different IOP measuring devices: Goldmann applanation tonometer (GAT), I-care pro, Tonopen XL, and Schiotz tonometr in patients who underwent DSEK. Methods: This was a prospective comparative study using a convenience cohort of post-DSEK patients with compact grafts. Post-DSEK patients had IOP measured using GAT, I-care Pro, Tonopen XL, and Schiotz tonometer. Measurements were compared and agreement assessed. Wilcoxon signed-rank test was used for comparison of means as variables did not show a normal distribution. Bland–Altman plots were used for assessing agreement. Results: Thirty eyes of 24 patients were included in the study. Mean time from DSEK surgery was 25.31 ± 13.05 months. Mean IOP with GAT, I-care pro, Tonopen XL, and Schiotz tonometer was 13.99 ± 3.76, 13.92 ± 3.36, 13.31 ± 3.89, and 12.83 ± 4.07, respectively. GAT, I-care pro, and Tonopen XL had similar mean IOP measurements (P = 0.135 and P = 0.551, respectively), while Schiotz tonometry measurements were higher (P = 0.046). Bland–Altman plots show good agreement between GAT, Tonopen XL, and I-care pro. GAT and Schiotz tonometry show less agreement, with large variations in the differences of measured IOP. Conclusions: IOP measurements in post-DSEK patients showed good agreement between GAT and either Tonopen XL or I-care pro. Schiotz tonometer has large variations in this patient group. IOP measurements and IOP difference between devices were not dependent on central corneal thickness.
机译:目的:Descemet剥离后的内皮角膜移植术(DSEK)患者容易发生眼内压(IOP)升高和青光眼。角膜特性以各种方式影响各种IOP测量设备。这项研究的目的是评估接受DSEK的四种不同IOP测量设备之间的一致性:Goldmann压平眼压计(GAT),I-care pro,Tonopen XL和Schiotz tonometr。方法:这是一项前瞻性比较研究,使用了DSEK术后紧凑型移植物的便利队列。 DSEK后患者使用GAT,I-care Pro,Tonopen XL和Schiotz眼压计测量眼压。比较测量值并评估一致性。由于变量未显示正态分布,因此使用了Wilcoxon符号秩检验来比较均值。用布兰德-奥特曼图来评估一致性。结果:24名患者的30只眼被纳入研究。 DSEK手术的平均时间为25.31±13.05个月。 GAT,I-care pro,Tonopen XL和Schiotz眼压计的平均IOP分别为13.99±3.76、13.92±3.36、13.31±3.89和12.83±4.07。 GAT,I-care pro和Tonopen XL的平均IOP测量值相似(分别为P = 0.135和P = 0.551),而Schiotz眼压测量值则更高(P = 0.046)。 Bland–Altman图显示了GAT,Tonopen XL和I-care pro之间的良好一致性。 GAT和Schiotz眼压测量法显示的一致性较差,但测得的IOP差异较大。结论:DSEK后患者的眼压测量结果显示,GAT与Tonopen XL或I-care pro的一致性良好。 Schiotz眼压计在该患者组中有很大的差异。设备之间的IOP测量值和IOP差异与中央角膜厚度无关。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号