Despite recent advances in implantable cardioverter defibrillator (ICD) technology, the long-term reliability of ICD leads remains a significant problem. Lead failures constitute a major risk for patients with an implantable cardioverter defibrillator. There is no clear consensus on treatment strategy of ICD lead failure and decision should be individualized. We report a pacing-dependent elderly male with ICD lead pacing failure secondary to insulation break resulting in recurrent syncope. We emphasize the technique of ''Add-on pacing lead implantation'' could be an effective and alternative to ICD lead replacement.
机译:尽管植入式心脏复律除颤器(ICD)技术最近取得了进步,但ICD导线的长期可靠性仍然是一个重大问题。铅故障对植入式心脏复律除颤器的患者构成重大风险。关于ICD导联失败的治疗策略尚无明确共识,应个体化决策。我们报告了起搏依赖的老年男性,继发于绝缘破坏导致ICD铅起搏失败,导致反复晕厥。我们强调“附加起搏引线植入”技术可能是替代ICD引线的有效替代方法。