首页> 外文期刊>Indian Journal of Dental Research >Effects of clinical use and sterilization on surface topography and surface roughness of three commonly used types of orthodontic archwires
【24h】

Effects of clinical use and sterilization on surface topography and surface roughness of three commonly used types of orthodontic archwires

机译:临床使用和灭菌对三种常用正畸弓丝表面形貌和表面粗糙度的影响

获取原文
           

摘要

Aim: To evaluate the changes in surface topography and roughness of stainless steel (SS), nickel-titanium and beta-titanium (β-Ti) archwires after clinical use and sterilization. Settings and Design: Thirty wires each of SS, nitinol, and β-Ti (3M Unitek) were tested in as received, as received and autoclaved, and clinically retrieved then autoclaved conditions. Materials and Methods: A sterilization protocol of 134°C for 18 min was performed using an autoclave. Surface topography of specimens from each subgroup was examined using an environmental scanning electron microscope (ESEM model Quanta 200, The Netherlands) at ×100, ×1000, and ×2500 magnifications. Surface roughness was measured using arithmetic mean roughness (Ra) values obtained from optical profilometric scanning (Taylor Hobson, Leicester, UK). Statistical Analysis: Data were analyzed by one-way analysis of variance and Tukey's post-hoc procedures. Results: Scanning electron microscope images revealed an increase in surface irregularities in SS and nitinol wires after clinical use. There was a significant increase in Ra values of SS orthodontic wires after intra-oral exposure (P = 0.0002). Conclusion: Surface roughness of SS wires increased significantly after clinical use. Autoclave sterilization did not affect considerably on surface characteristics of any archwire.
机译:目的:评估临床使用和灭菌后不锈钢(SS),镍钛和β-钛(β-Ti)弓丝表面形貌和粗糙度的变化。设置和设计:分别对30根SS,镍钛诺和β-Ti(3M Unitek)的电线进行接收,接收和高压灭菌测试,然后临床恢复高压灭菌条件。材料和方法:使用高压釜进行134°C的18分钟灭菌方案。使用环境扫描电子显微镜(ESEM模型Quanta 200,荷兰)以×100,×1000和×2500的放大倍数检查每个亚组样品的表面形貌。使用从光学轮廓仪扫描获得的算术平均粗糙度(Ra)值(Taylor Hobson,英国莱斯特)测量表面粗糙度。统计分析:通过单因素方差分析和Tukey事后程序对数据进行分析。结果:扫描电子显微镜图像显示,临床使用后SS和镍钛诺丝表面不规则性增加。口腔内暴露后,SS正畸线的Ra值显着增加(P = 0.0002)。结论:不锈钢丝的表面粗糙度在临床使用后明显增加。高压灭菌器对任何弓丝的表面特性没有明显影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号