首页> 外文期刊>Analytical Sciences >Multielement Determination of Trace Metals in River Water (Certified Reference Material, JSAC 0301-1) by High Efficiency Nebulization ICP-MS after 100-fold Preconcentration with a Chelating Resin-Packed Minicolumn
【24h】

Multielement Determination of Trace Metals in River Water (Certified Reference Material, JSAC 0301-1) by High Efficiency Nebulization ICP-MS after 100-fold Preconcentration with a Chelating Resin-Packed Minicolumn

机译:螯合树脂填充小柱预浓缩100倍后,通过高效雾化ICP-MS法测定河流水中的痕量金属(认证参考材料,JSAC 0301-1)

获取原文
           

摘要

The determination of 34 trace metals in a river water certified reference material (CRM), i.e. JSAC 0301-1, which was issued by the Japan Society for Analytical Chemistry in January 2004, was performed by ICP-MS with a high efficiency nebulizer after preconcentration with a laboratory-made chelating resin-packed minicolumn, with which trace metals were concentrated 100-fold from 50 mL of a river water sample to 0.5 mL of the final analysis solution. Trace metals in JSAC 0301-1 were observed in the concentration range from 19 µg L-1 of Al to 0.000053 µg L-1 of Bi. It was found that most of the concentrations of trace metals, including rare earth elements (REEs), in JSAC 0301-1 were lower than those in JAC 0031, which was also a previously issued CRM prepared with water from the same river as that of JSAC 0301-1. The low concentrations of trace metals in JSAC 0301-1 might be attributed to the fact that there was a heavy rain before collecting the original water sample to prepare the present CRM. Furthermore, the REE distribution patterns of JSAC 0301-1, JAC 0031 and the average values of river water samples in Japan were parallel to each other. These results indicate that the distributions of REEs in JSAC 0301-1 and JAC 0031 were the typical ones of river water samples in Japan.
机译:2004年1月,由日本分析化学学会发布的JSAC 0301-1,即河水认证参考材料(CRM)中的34种痕量金属的测定是通过ICP-MS在预浓缩后通过高效雾化器进行的用实验室制得的填充螯合树脂的微型柱,微量金属从50 mL的河水样品浓缩到0.5 mL的最终分析溶液,浓缩了100倍。观察到JSAC 0301-1中的微量金属的浓度范围为19 µg L-1的Al至0.000053 µg L-1的Bi。发现JSAC 0301-1中的大多数痕量金属,包括稀土元素(REE)的浓度都低于JAC 0031中的含量,JAC 0031也是先前发布的CRM,它是使用与JSAC 0301-1。 JSAC 0301-1中痕量金属的浓度低可能是由于在收集原始水样以准备本CRM之前有大雨。此外,日本的JSAC 0301-1,JAC 0031的REE分布模式与河流水样的平均值相互平行。这些结果表明,JSAC 0301-1和JAC 0031中的REE分布是日本典型的河水样品。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号