Improvement of an Automatic HPLC System for Nitropolycyclic Aromatic Hydrocarbons: Removal of an Interfering Peak and Increase in the Number of Analytes

机译:硝基多环芳烃自动HPLC系统的改进:干扰峰的去除和分析物数量的增加

获取原文
获取外文期刊封面目录资料

摘要

An automatic HPLC system for analyzing nitropolycyclic aromatic hydrocarbons (NPAHs, nitroarenes) in airborne particulates was previously described (Anal. Chim. Acta, 2001, 445, 20). Some problems with this system were that in generated a peak originating from an ascorbic acid solution that elutes at a retention time close to that of 1,6-dinitropyrene (DNP), and that it was able to analyze only 1,3-, 1,6-, 1,8-DNPs and 1-nitropyrene (1-NP). Here, we describe an improved system that effectively removes the interfering peak by introducing an ODS column just after the pump for the ascorbic acid solution, and which is capable of analyzing several additional compounds (2-, 4-NPs, 2-nitrofluorene, 6-nitrochrysene, 7-nitrobenz[a]anthracene, 3-nitroperylene and 6-nitrobenzo[a]pyrene etc.). The improved sensitivities were achieved by concentrating the compounds in a benzene-ethanol extract from airborne particulates, by increasing the loading time of the sample solution from 20 to 38 min, and by increasing the flow rate of an ascorbic acid solution from 1.3 to 1.8 mL/min.
机译:先前已经描述了一种自动HPLC系统,用于分析空气中颗粒中的硝基多环芳烃(NPAHs,硝基芳烃)(Anal。Chim。Acta,2001,445,20)。该系统存在的一些问题是,它会在抗坏血酸溶液中产生一个峰,该峰在保留时间接近1,6-二硝基py(DNP)的时候洗脱下来,并且只能分析1,3-,1。 ,6-,1,8-DNP和1-硝基py(1-NP)。在这里,我们描述了一种经过改进的系统,该系统可以在泵送抗坏血酸溶液之后立即引入ODS色谱柱,从而有效地去除干扰峰,并且该色谱柱能够分析多种其他化合物(2-,4-NP,2-硝基芴,6 -硝基氯,7-硝基苯并[a]蒽,3-硝基per和6-硝基苯并[a] py等)。通过从空气中的颗粒物的苯乙醇提取物中浓缩化合物,将样品溶液的上样时间从20分钟增加到38分钟,以及将抗坏血酸溶液的流速从1.3毫升增加到1.8毫升,可以实现更高的灵敏度。 /分钟。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号