Duplicate analysis method: a cheaper alternative to commercial IQC materials in limited resource settings for monitoring CD4 testing

机译:重复分析方法:在有限的资源设置中以便宜的方式替代商业IQC材料,以监控CD4测试

获取原文
           

摘要

Background India has a large number of HIV infected patients being followed up at anti-retroviral therapy (ART) centers. The patients are regularly offered CD4 count estimation for deciding their eligibility for ART initiation as well as for monitoring response to ART, making CD4 count estimation a very critical test. Hence, quality control of CD4 testing is utmost important for ultimate success of ART program. As the commercial controls are very expensive, internal quality control (IQC), at present, is being done by duplicate analysis method using previous day samples in most of the laboratories. Hence the study was undertaken to review performance of duplicate analysis method for monitoring daily IQC. Methods Quality control (QC) data from 11 Indian laboratories using duplicate analysis and/or commercial controls for IQC of CD4 testing was collected for reviewing information on QC parameters such as precision, accuracy and trend monitoring. Precision was determined by r 2 values and mean % variation for duplicate analysis and coefficient of variation (% CV) for commercial controls. Accuracy was monitored by rate of QC failures for both the types of control and trend monitoring was done by plotting LJ charts for commercial controls and by plotting daily % variation for duplicate analysis. Results The laboratories using duplicate analysis for IQC showed good precision with mean % variation ranging from 0.5 to 7.2. There was good match between r 2 values and % CV of the laboratories performing both the types of QC methods. Rates of QC failures were 2.3 for duplicate analysis and 3 per laboratory-year for IMMUNO-TROL controls. Daily trend monitoring showed fluctuation of daily counts around mean in LJ charts and of percent variation around 0% in duplicate analysis method. Commercially available controls showed limitations such as altered specimen quality leading to difficulties in manual gating and issues with the establishment of laboratory range. Conclusion Duplicate analysis can serve as a cheaper alternative to commercially available controls for IQC of CD4 testing especially when supplemented with other QC measures for controlling variations caused by reagent, equipment, staff and environment in addition to the successful participation in External Quality Assurance programme.
机译:背景技术印度有大量的HIV感染患者正在抗逆转录病毒治疗(ART)中心接受随访。定期为患者提供CD4计数估算,以决定他们是否接受抗逆转录病毒治疗以及监测对ART的反应,从而使CD4计数估算成为非常关键的测试。因此,CD4测试的质量控制对于ART程序的最终成功至关重要。由于商业控制非常昂贵,因此目前,大多数实验室中都使用前一天的样品通过重复分析方法进行内部质量控制(IQC)。因此,本研究旨在审查重复分析方法在监测日常IQC方面的性能。方法收集来自印度11个实验室的质量控制(QC)数据,该数据使用重复分析和/或用于CD4测试IQC的商业对照进行收集,以审查有关QC参数的信息,例如精度,准确性和趋势监测。精密度由r 2 值和重复分析的均值变异百分比确定,而商用控件的变异系数(%CV)确定。通过控制类型的QC失败率来监视准确性,并通过绘制LJ图进行商业控件并绘制每日%变化用于重复分析来进行趋势监视。结果对IQC进行重复分析的实验室显示出良好的精度,平均%变异范围为0.5至7.2。 r 2 值与执行两种QC方法的实验室的%CV值之间具有良好的匹配。重复分析的QC失败率为2.3,IMMUNO-TROL对照的每个实验室年为3。每日趋势监控显示,LJ图表中的每日计数围绕均值波动,而重复分析方法中的每日变化百分比约为0%。市售的对照品显示出局限性,例如改变标本质量导致手动选通困难以及建立实验室范围的问题。结论重复分析可以比CD4检测的IQC的市售对照便宜得多,尤其是在成功参加外部质量保证计划的同时,补充其他QC措施以控制由试剂,设备,人员和环境引起的变异时,重复分析可以作为便宜的替代方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号