首页> 外文期刊>African Vision and Eye Health >A comparison of intraocular pressure values obtained with the Tono-Pachymeter NT530P, iCare? rebound tonometer and Goldmann applanation tonometer
【24h】

A comparison of intraocular pressure values obtained with the Tono-Pachymeter NT530P, iCare? rebound tonometer and Goldmann applanation tonometer

机译:用Tono-Pachymeter NT530P,iCare?获得的眼压值的比较。回弹眼压计和戈德曼压平眼压计

获取原文
   

获取外文期刊封面封底 >>

       

摘要

The purpose of this study was to compare the intraocular pressure (IOP) values measured with the Tono-Pachymeter NT530P (Tonopachy?) and the iCare? rebound tonometer (iCare?) with those obtained by the Goldmann applanation tonometer (GAT). The right eyes of 105 subjects aged 18 to 82 years (mean age = 29.27 ± 14.67 years) were assessed with the three tonometers. Central corneal thickness (CCT) was measured first using the Tonopachy? and then IOP was measured by Tonopachy?, iCare? and GAT. The data was analyzed with descriptive statistics, paired? t-test, correlation and regression analysis. The Bland-Altman method of analysis was used to evaluate agreements between the sets of data from the three devices. The CCT values ranged from 440 μm to 606 μm (mean= 518.49 ± 33.01 μm). There was little or no correlation between CCT and IOP for any of the instruments used in this study (r = 0.29 for Tonopachy?, r = 0.22 for iCare?, r = 0.17 for GAT). The mean IOP measured with the Tonopachy? was 14.31 ± 3.57 mmHg (range 8.7 mmHg to 31 mmHg) and 16.64 ± 4.38 mmHg (range 8 mmHg to 32 mmHg) using the iCare?. The mean IOP measured with the GAT was 14.79 ± 3.09 mmHg (range 8.7 mmHg to 29.7 mmHg). Using the Bland-Altman method, the upper and lower limits of agreement between the Tonopachy? and GAT, iCare? and GAT, iCare? and Tonopachy? were 5.1 mmHg and –4.2 mmHg, 8.6 mmHg and –4.9 mmHg, 7.5 mmHg and –2.8 mmHg respectively. In 79.1% of the eyes studied, the mean IOP difference between Tonopachy? and GAT was less than 3 mmHg and in 20.9% of the eyes, the difference was greater than 3 mmHg. However, mean IOP differences of greater than 3 mmHg were obtained by iCare? in comparison with GAT (40%) and Tonopachy? (34.3%) respectively. Findings of this study suggest that the Tonopachy? yielded IOP readings that were consistent with those of GAT values while iCare? yielded higher IOP values compared to both GAT and Tonopachy?. (S Afr Optom 2011 70(3) 109-116)
机译:这项研究的目的是比较用Tono-Pachymeter NT530P(Tonopachy?)和iCare?测量的眼内压(IOP)值。回弹眼压计(iCare?)和高曼压平眼压计(GAT)获得的值。用三个眼压计评估了105位年龄在18至82岁(平均年龄= 29.27±14.67岁)的受试者的右眼。首先使用Tonopachy?来测量中央角膜厚度(CCT)。然后用Tonopachy?,iCare?测量IOP。和GAT。使用描述性统计数据对数据进行分析,配对吗? t检验,相关和回归分析。 Bland-Altman分析方法用于评估来自三个设备的数据集之间的一致性。 CCT值的范围为440μm至606μm(平均值= 518.49±33.01μm)。对于本研究中使用的任何仪器,CCT和IOP之间几乎没有相关性(对于Tonopachy™,r = 0.29,对于iCare™,r = 0.22,对于GAT,r = 0.17)。用Tonopachy测得的平均IOP?使用iCare?测得的结果为14.31±3.57 mmHg(范围为8.7 mmHg至31 mmHg)和16.64±4.38 mmHg(范围为8 mmHg至32 mmHg)。用GAT测得的平均IOP为14.79±3.09 mmHg(范围为8.7 mmHg至29.7 mmHg)。使用Bland-Altman方法,Tonopachy之间的一致性上限和下限?和GAT,iCare?和GAT,iCare?和Tonopachy?分别为5.1 mmHg和–4.2 mmHg,–8.6 mmHg和–4.9 mmHg,–7.5 mmHg和–2.8 mmHg。在研究的79.1%的眼睛中,Tonopachy?而GAT小于3 mmHg,在20.9%的眼睛中,差异大于3 mmHg。但是,通过iCare?获得的平均IOP差异大于3 mmHg。与GAT(40%)和Tonopachy相比? (34.3%)。这项研究的结果表明,Tonopachy?在iCare期间产生的IOP读数与GAT值一致。与GAT和Tonopachy?相比,IOP值更高。 (S Afr Optom 2011 70(3)109-116)

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号