Phantom-based quality assurance measurements in B-mode ultrasound:

机译:B模式超声中基于幻影的质量保证测量:

获取原文
       

摘要

Recommended phantom-based quality assurance measurements in B-mode ultrasound (US) may be tedious. For the purpose of cost-effective US quality assurance it is important to evaluate measurements that effectively reflect the quality of US scanner. To find out which recommended phantom-based quality assurance measurements are effective in detecting dead or weak transducer elements or channels in US scanners when visual image analysis and manual measurements are used. Altogether 66 transducers from 33 US scanners were measured using a general purpose phantom and a transducer tester. The measurements were divided into two groups. Group I consisted of phantom-based uniformity measurement, imaging the air with a clean transducer (air image) and measuring the transducer with the transducer tester, and group II of phantom-based measurements of depth of penetration, beam profile, near field, axial and lateral resolution, and vertical and horizontal distance accuracy. The group II measurements were compared to group I measurements. With group I measurements, the results with 20% of the transducers were found defective. With 35% of the transducers the results were considered defective in group II measurements. Concurrent flaws in both groups were found with 11% of the transducers. Phantom-based measurements of depth of penetration, beam profile, near field, axial and lateral resolution, and vertical and horizontal distance accuracy did not consistently detect dead or weak transducer elements or channels in US scanners.
机译:建议在B模式超声(US)中基于幻像的质量保证测量可能很乏味。为了具有成本效益的美国质量保证,评估有效反映美国扫描仪质量的测量非常重要。为了找出在使用可视图像分析和手动测量时,哪些推荐的基于幻像的质量保证测量有效检测US扫描仪中的死角或弱换能器元件或通道。使用通用体模和换能器测试仪对来自33个美国扫描仪的66个换能器进行了测量。测量分为两组。第一组包括基于幻像的均匀性测量,使用清洁的换能器对空气成像(空气图像)并使用换能器测试仪测量换能器,第二组包括对穿透深度,光束轮廓,近场,轴向的基于幻像的测量和横向分辨率,以及垂直和水平距离的精度。将II组的测量结果与I组的测量结果进行比较。在进行I组测量时,发现20%的传感器的结果有缺陷。对于35%的换能器,结果在II组测量中被认为是有缺陷的。两组中有11%的换能器同时出现缺陷。基于透视图的穿透深度,光束轮廓,近场,轴向和横向分辨率以及垂直和水平距离精度的测量无法始终检测到US扫描仪中的死角或弱换能器元件或通道。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号