Irreversible electroporation (IRE) is an ablation technique that is being investigated as a potential treatment of pancreatic cancer. However, many of these patients have a metallic stent in the bile duct, which is recognized as a contraindication for IRE ablation. We report a case in which an IRE ablation in the region of the pancreatic head was performed on a patient with a metallic stent which led to severe complications. After the treatment, the patient suffered from several complications including perforation of the duodenum and transverse colon, and bleeding from a branch of the superior mesenteric artery which eventually lead to her death. Therefore, we believe it is important to be aware of this if an IRE ablation close to a metallic stent is considered.
机译:不可逆电穿孔(IRE)是一种消融技术,正在研究作为一种潜在的胰腺癌治疗方法。但是,这些患者中有许多在胆管中有金属支架,这被认为是IRE消融的禁忌证。我们报道了一种情况,其中在带有金属支架的患者上进行了胰头区域的IRE消融,这导致了严重的并发症。治疗后,患者患有多种并发症,包括十二指肠和横结肠穿孔以及肠系膜上动脉分支出血,最终导致死亡。因此,我们认为,重要的是要考虑考虑在金属支架附近进行IRE消融。