Functional magnetic resonance imaging (fMRI) is one of the techniques capable of detecting human brain function noninvasively. Although many fMRI studies have adopted echo-time (TE) of 30 to 40ms aiming to maximize blood oxygen level-dependent (BOLD) signal change, large distortion and signal dropout occur near air-filled spaces such as frontal sinus in functional imaging because of the magnetic field irregularities. Theoretically, shorter TE could decrease the magnetic distortion and signal dropout near air. However, it has not been determined whether short TE of less than 30ms may affect the statistical t value of functional image, which is the most important factor to evaluate results of fMRI study. In the present study, we assessed whether shorter TE can be utilized for fMRI study of the human brain. Nineteen healthy volunteers were scanned using fMRI during finger tapping and viewing checker-circle tasks. Signal dropout near the sinus, statistical t value, percent signal change, and modified temporal signal to noise ratio (mtSNR) of functional images were evaluated in the primary motor cortex and primary visual cortex while TE was varied between 20 and 35ms. Signal dropout was the smallest at a TE of 20ms, whereas there was no significant effect of TE on the statistical t value both in the primary motor cortex and in the primary visual cortex activity. These findings suggest that 20ms is the optimal TE value in the range of 20ms to 35ms for fMRI.
机译:功能磁共振成像(fMRI)是能够无创地检测人脑功能的技术之一。尽管许多功能磁共振成像研究已采用30到40毫秒的回波时间(TE),旨在最大程度地提高血氧水平相关(BOLD)信号的变化,但由于功能性成像,在充满空气的空间(如额窦)附近会出现较大的失真和信号丢失磁场不规则。从理论上讲,较短的TE可以减少空气附近的磁畸变和信号丢失。然而,尚未确定小于30ms的短TE是否会影响功能图像的统计值,这是评估fMRI研究结果的最重要因素。在本研究中,我们评估了较短的TE是否可以用于人脑的fMRI研究。在手指敲击和查看棋盘圈任务期间,使用fMRI对19名健康志愿者进行了扫描。在TE在20到35ms之间变化的同时,评估了运动窦和初级视觉皮质中窦道附近的信号丢失,统计t值,百分比信号变化以及功能图像的修改时间信噪比(mtSNR)。在20ms的TE时信号丢失最小,而TE对初级运动皮层和初级视觉皮层活动的统计t值均无显着影响。这些发现表明,对于功能磁共振成像,20ms是在20ms至35ms范围内的最佳TE值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号