Errata for the paper entitled “Effects of Eccentric Impeller Position on Radial Passive Stability in a Magnetically Levitated Centrifugal Blood Pump with a Double Volute” [Advanced Biomedical Engineering Vol. 7 (2018) pp.63-71]

机译:题为“偏心叶轮位置对带有双蜗壳的磁悬浮离心血泵中径向被动稳定性的影响”的论文勘误表。 7(2018)pp.63-71]

获取原文
       

摘要

Upon double-checking the experimental setup, the authors discovered that the impeller diameter and radial clearance of the prototype geometry used in the experiment were incorrect. The correct diameter and radial clearance of prototype geometry were 51 mm and 1.5 mm, respectively, although a diameter of 52 mm and radial clearance of 1.0 mm were described in the original paper and these dimensions were used in some calculations. However, inaccuracy in the description of the experimental setting does not affect the results of computational fluid dynamics (CFD) analysis.
机译:在仔细检查实验设置后,作者发现实验中使用的原型几何形状的叶轮直径和径向间隙不正确。原型几何的正确直径和径向间隙分别为51毫米和1.5毫米,尽管在原始论文中描述了52毫米的直径和1.0毫米的径向间隙,并且这些尺寸已用于某些计算中。但是,实验设置描述中的不准确性不会影响计算流体动力学(CFD)分析的结果。