首页> 外文期刊>Computer Assisted Surgery >A novel device for preoperative registration and automatic tracking in cranio-maxillofacial image guided surgery
【24h】

A novel device for preoperative registration and automatic tracking in cranio-maxillofacial image guided surgery

机译:颅颌面图像引导手术中术前套准和自动跟踪的新型装置

获取原文
           

摘要

Objective: Two key issues in image guided surgery are accurate patient-to-image registration and ongoing tracking of the patient's motion. To address these concerns, a novel device for preoperative registration and automatic tracking was designed, and the accuracy attainable with the device was evaluated in experiments with a skull and in a clinical study.Methods: The device consists of a system of four titanium screws and four fluorescent spheres ?xed to carbon bars which can be easily mounted on the maxillary dentition splint. Before surgery, CT image data of a skull with the device in place was acquired and registered in a navigation system. The rigidity and reproducibility of positioning of the device were measured in 15 repeated CT acquisitions of the skull with the device in place. The registration accuracy was compared to that obtained using micro-screw markers ?xed to the maxillary alveolus. To determine the potential of the device in aiding image guided cranio-maxillofacial surgery, registration accuracy and surgical outcome were assessed.Results: Fifteen tests were performed for CT scanning with no loosening of the splint and device. The arithmetic mean of the standard deviation (SD) ranged from 0.47?mm to 0.70?mm. When the device was used for registration, the mean deviations for the eight anatomical structures investigated ranged from 0.56?mm at the left infra-orbital foramen to 0.96?mm at the right temple. Compared with the method in which titanium screws are fixed to the maxillary alveolus, the target registration error (TRE) obtained using the new device was much less. Using this device, clinical reduction of a zygomatic-orbital-maxillary complex fracture was successfully completed with a registration discrepancy of less than 0.5?mm.Conclusions: By successfully addressing the two key issues of image guided surgery, the device could be considered accurate and potentially useful for assisting in cranio-maxillofacial surgery.
机译:目的:图像引导手术中的两个关键问题是准确的患者到图像配准以及对患者运动的持续跟踪。为了解决这些问题,设计了一种用于术前套准和自动跟踪的新型设备,并在颅骨实验和临床研究中评估了该设备可达到的准确性。方法:该设备由四个钛合金螺钉和四个固定在碳棒上的荧光球可以很容易地安装在上颌牙列夹板上。手术前,在装置已就位的情况下获取颅骨的CT图像数据,并将其记录在导航系统中。在放置了该装置的15次重复CT颅骨CT采集中,测量了该装置定位的刚度和可重复性。将套准精度与使用固定在上颌牙槽上的微螺钉标记获得的套准精度进行比较。为了确定该装置在影像引导下的颅颌面外科手术中的潜力,评估了套准准确性和手术结果。结果:进行了15次CT扫描测试,未松动夹板和装置。标准偏差(SD)的算术平均值范围为0.47?mm至0.70?mm。当使用该装置进行配准时,所研究的八个解剖结构的平均偏差范围从左眼眶下孔的0.56?mm到右太阳穴处的0.96?mm。与将钛螺钉固定在上颌牙槽上的方法相比,使用新设备获得的目标配准误差(TRE)小得多。使用该装置,成功完成了reduction眶眶上颌复杂骨折的临床复位,配准差异小于0.5?mm。结论:通过成功解决影像引导手术的两个关键问题,该装置可被认为是准确且可能对协助颅颌面手术有用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号