首页> 外文期刊>Circulation journal >Impact of Pacing and High-Pass Filter Settings on Ventricular Bipolar Electrograms in Implantable Cardioverter Defibrillator Systems – Implication of Predictors for Inappropriate Therapy Caused by Oversensing of Repolarization Electrograms –
【24h】

Impact of Pacing and High-Pass Filter Settings on Ventricular Bipolar Electrograms in Implantable Cardioverter Defibrillator Systems – Implication of Predictors for Inappropriate Therapy Caused by Oversensing of Repolarization Electrograms –

机译:起搏和高通滤波器设置对植入式心脏复律除颤器系统中的心室双极电图的影响–对复极化电图的过度感应导致治疗不当的预测因素的涵义–

获取原文
           

摘要

Background: Predictors of T wave oversensing with implantable cardioverter-defibrillator (ICD) systems remains to be clarified. Methods and Results: Thirteen consecutive patients who underwent ICD implantations were included. The depolarization (R) and repolarization (T) of bipolar electrograms during baseline, AAI and DDD modes, and an isoproterenol (ISO) infusion were evaluated. The R wave amplitude during DDD was significantly lower as compared to that during the other conditions in all high-pass filter settings. In contrast, there was no significant difference in the T wave amplitude during the DDD as compared to the other conditions. With the DDD, there was a significantly higher incidence of a T/R ratio of greater than 0.25 as compared to that with the other conditions. T wave amplitude in Brugada syndrome was significantly higher than that in non-Brugada syndrome. The existence of Brugada syndrome and T/R ratio during the AAI with a high-pass filter setting of 10/20Hz was an excellent predictor of T wave oversensing in the follow-up period. Conclusions: DDD had a significant impact on the R wave amplitude reduction and the T/R ratio during AAI can be predictors of T wave oversensing. These findings have important implications for inappropriate shocks due to T wave oversensing. ( Circ J 2011; 75: 2095-2104)
机译:背景:植入式心脏复律除颤器(ICD)系统对T波过度感应的预测仍有待澄清。方法和结果:纳入了13例接受ICD植入的患者。评估了基线,AAI和DDD模式以及异丙肾上腺素(ISO)输注期间双极电描记图的去极化(R)和复极化(T)。在所有高通滤波器设置中,DDD期间的R波幅值均比其他条件下的低。相反,与其他条件相比,DDD期间的T波振幅没有显着差异。使用DDD时,与其他条件相比,T / R比率大于0.25的发生率显着更高。 Brugada综合征的T波振幅明显高于非Brugada综合征。高通滤波器设置为10 / 20Hz的AAI期间Brugada综合征的存在和T / R比率是随访期间T波过度感应的良好预测指标。结论:DDD对R波振幅的减小有显着影响,而AAI期间的T / R比可以预测T波的过度感应。这些发现对于由于T波过感应而引起的不适当冲击具有重要意义。 (Circ J 2011; 75:2095-2104)

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号