首页> 外文期刊>Bangladesh Journal of Medical Physics >Responses of LiF Thermoluminescence Dosimeters to Diagnostic 60Co Teletherapy Beams
【24h】

Responses of LiF Thermoluminescence Dosimeters to Diagnostic 60Co Teletherapy Beams

机译:LiF热致发光剂量计对60Co诊断远距放射束的响应

获取原文
       

摘要

Thermoluminescence Dosimeter (TLD) is used as an ionizing radiation detector in the field of clinical radiotherapy, diagnostic radiology, personal radiation monitoring as well as in environmental radiation dosimetry. The sensitivity of all the TLDs is not the same even though the manufacturer produced them in the same batch. The response of TLD at various doses is also not always linear. In the present study fifty-seven rod shaped LiF TLDs had been irradiated by doses of 50cGy, 100cGy, 150cGy and 200cGy using 60Co teletherapy unit of Dhaka Medical College (Alcyon II, CGR, McV, France). The TLDs were placed in the grooves of a slab phantom and then covered by a 0.5cm thick plexiglass sheet. The phantom was then irradiated in the reference condition (10x10cm2 field size, 80cm SSD). After irradiation, readout of TLDs was done using a Harshaw TLD reader (Model 3500). Measurements were performed three times for each dose value and their average was taken. For convenience all fifty-seven TLDs were divided into nineteen groups. Grouping was done according to the response of the TLDs for a fixed dose value. In a particular group of TLDs, variation of the response of the dosimeters is the smallest one. The dose (cGy) – response (nC) linearity curve of TLDs for each of the nineteen groups showed to be very close to unity. Bangladesh Journal of Medical Physics Vol.8 No.1 2015 14-21
机译:热致发光剂量计(TLD)在临床放射治疗,放射诊断学,个人放射监测以及环境放射剂量测定领域中用作电离放射检测器。即使制造商以同一批次生产它们,所有TLD的灵敏度也不相同。 TLD在各种剂量下的响应也不总是线性的。在本研究中,使用达卡医学院的 60 Co远程治疗装置(Alcyon II,CGR,McV,法国)以50cGy,100cGy,150cGy和200cGy的剂量辐照了57个棒状LiF TLD。 。将TLD放置在平板模型的凹槽中,然后用0.5cm厚的有机玻璃板覆盖。然后以参考条件(10x10cm 2 场尺寸,80cm SSD)照射体模。辐照后,使用Harshaw TLD阅读器(型号3500)读取TLD。每个剂量值进行三次测量,取其平均值。为方便起见,将全部57个TLD分为19个组。根据TLD对固定剂量值的响应进行分组。在一组特定的TLD中,剂量计响应的变化是最小的。对于19个组中的每个组,TLD的剂量(cGy)–响应(nC)线性曲线均显示非常接近统一。孟加拉国医学物理学杂志2015年第8卷第1期14-21

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号