Losartan potassium is an antihypertensive non-peptide agent, which exerts its action by specific blockade of angiotensin II receptors. The aim of the present study was the validation and application of analytical methods for the quality control of losartan potassium 50 mg in pharmaceutical capsules, using direct and first-derivative UV spectrophotometry. Based on losartan potassium spectrophotometric characteristics, a signal at 205 nm of the zero-order spectrum and a signal at 234 nm of the first-derivative spectrum, were found adequate for quantification. The results were used to compare these instrumental techniques. The linearity between the signals and concentrations of losartan potassium in the ranges of 3.0-7.0 mg L-1 and 6.0-14.0 mg L-1 for direct and first-derivative spectrophotometry in aqueous solutions, respectively, presented a correlation coefficient (r) of 0.9999 in both cases. The methods were applied for losartan potassium in capsule dosage obtained from local pharmacies, and were shown to be efficient, easy to apply and low cost. These methods do not use polluting reagents and require relatively inexpensive equipment.
机译:氯沙坦钾是一种抗高血压的非肽类药物,通过特异性阻断血管紧张素II受体发挥其作用。本研究的目的是使用直接和一阶导数紫外分光光度法验证并验证用于药物胶囊中50 mg氯沙坦钾质量控制的分析方法。基于氯沙坦钾分光光度法的特性,发现零级光谱在205 nm处的信号和一阶导数光谱在234 nm处的信号足以定量。结果用于比较这些仪器技术。直接和一阶导数分光光度法在水溶液中的信号和氯沙坦钾浓度在3.0-7.0 mg L-1和6.0-14.0 mg L-1范围之间的线性关系,相关系数(r)为两种情况下均为0.9999。该方法应用于从当地药房获得的氯沙坦钾胶囊剂量,经证实是有效,易于使用且成本低廉的。这些方法不使用污染试剂,并且需要相对便宜的设备。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号