Comparison of sodium ion levels between an arterial blood gas analyzer and an autoanalyzer in preterm infants admitted to the neonatal intensive care unit: a retrospective study

机译:新生儿重症监护病房收治的早产儿动脉血气分析仪和自动分析仪钠离子水平的比较:一项回顾性研究

获取原文
       

摘要

Background The difference in sodium ion levels determined with direct and indirect methods often exceeds the permissible limit clinically. Additionally, no previous study has assessed the difference in the sodium ion levels between direct and indirect methods in premature infants. Therefore, the present study aimed to compare sodium ion levels obtained using an arterial blood gas analyzer (ABGA; direct method) and an autoanalyzer (indirect method) to determine whether they are equivalent in premature infants. Methods The present retrospective study included 450 preterm infants (weight, Results The mean sodium ion levels were 134.6?±?3.5?mmol/L using the ABGA and 138.8?±?4.7?mmol/L using the autoanalyzer ( P 4?mmol/L between the two methods (odds ratio?=?2.870, P Conclusion The sodium ion levels determined using the ABGA and the autoanalyzer might not be equivalent in premature infants admitted to the NICU. Therefore, clinicians should be careful when diagnosing sodium ion imbalance in premature infants and providing treatment.
机译:背景技术直接法和间接法测定的钠离子水平差异通常在临床上超过允许的范围。另外,以前的研究还没有评估早产儿直接和间接方法之间钠离子水平的差异。因此,本研究旨在比较使用动脉血气分析仪(ABGA;直接方法)和自动分析仪(间接方法)获得的钠离子水平,以确定它们在早产儿中是否相等。方法这项回顾性研究包括450名早产儿(体重,结果),使用ABGA分析仪的平均钠离子水平为134.6?±?3.5?mmol / L,使用自动分析仪的平均钠离子水平为138.8?±?4.7?mmol / L(P 4?mmol /两种方法之间的差异为L(奇数比≤2.870,P)结论使用ABGA和自动分析仪测定的钠离子水平对于新生儿重症监护病房(NICU)的早产儿可能不相等,因此,临床医生在诊断食管癌中钠离子失衡时应谨慎。早产儿并提供治疗。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号