Bone density: comparative evaluation of Hounsfield units in multislice and cone-beam computed tomography

机译:骨密度:多层和锥束计算机断层扫描中Hounsfield单位的比较评估

获取原文
           

摘要

The aim of this study was to evaluate the validity of the bone density value of potential implant sites in HU obtained by a specific cone-beam computed tomography (CBCT) device. In this study, the HU values obtained using a MSCT scanner were used as the gold standard. Twenty mandibles (40 potential implant sites) were scanned using an MSCT scanner (Somatom Sensation 40) and a CBCT scanner (i-CAT). The MSCT images were evaluated using the Syngo CT Workplace software and the CBCT images, using the XoranCat software. The images were evaluated twice by three oral radiologists, at 60 day intervals. The trabecular bone density of the same area was evaluated on both images. Intraclass coefficients (ICC) were calculated to examine the agreement between the examiners and between the two periods of evaluation. The bone density and area of the ROI were compared by the Student t test and Bland-Altman analysis. ICCs were excellent. The mean HU value obtained using CBCT (418.06) was higher than that obtained using MSCT (313.13), with a statistically significant difference (p < 0.0001). In addition, Bland-Altman analysis showed that the HU measures were not equivalent. In conclusion, the bone density in HU with CBCT images obtained using the device studied proved unreliable, since it was higher than that obtained using MSCT.
机译:这项研究的目的是评估通过特定的锥形束计算机断层扫描(CBCT)设备获得的HU潜在植入部位的骨密度值的有效性。在这项研究中,使用MSCT扫描仪获得的HU值被用作黄金标准。使用MSCT扫描仪(Somatom Sensation 40)和CBCT扫描仪(i-CAT)扫描了20个下颌骨(40个潜在的植入部位)。使用Syngo CT Workplace软件评估MSCT图像,使用XoranCat软件评估CBCT图像。由三名口腔放射科医生以60天为间隔对图像进行两次评估。在两个图像上都评估了同一区域的小梁骨密度。计算类内系数(ICC)以检查考官之间以及两个评估期间之间的一致性。通过Student t检验和Bland-Altman分析比较了ROI的骨密度和面积。 ICC非常出色。使用CBCT(418.06)获得的平均HU值高于使用MSCT(313.13)获得的平均HU值,具有统计学显着性差异(p <0.0001)。此外,Bland-Altman分析表明,HU量度不相等。总之,使用所研究的设备获得的具有CBCT图像的HU中的骨密度被证明是不可靠的,因为它比使用MSCT获得的更高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号