The comparison of erythrocyte sedimentation rate (ESR) modify Westergren Caretium Xc-A30 and Westergren Manual in Clinical Pathology Laboratory, Sanglah General Hospital, Denpasar, Bali

机译:巴厘岛登巴萨桑格拉总医院临床病理实验室的Westergren Caretium Xc-A30和Westergren手册中的红细胞沉降率(ESR)的比较

获取原文
           

摘要

Background: The erythrocyte sedimentation rate (ESR) test is one of the most widely performed laboratory test to assess acute phase of inflammation, infection, autoimmune or malignancy. The reference method for ESR measurement was introduced by Westergren. This method is generally easy, un-expensive, yet, time consuming. Automatic instrument has been introduced to decrease the measurement time. Caretium XC-A30 is automatic instrument to measure ESR; it is used in Sanglah General Hospital Denpasar without previous research. Aim of this study was to assess comparison, correlation and agreement of Caretium XC- A30 to the reference method. Methods: A total of 35 samples in consecutive sampling were taken during August 2018 at Sanglah General Hospital, Denpasar. The samples were examined for ESR with Caretium XC-A30 and manual Westergren. Comparison, correlation and agreement were tested for both methods used. Results: The Westergren method was the reference method and Caretium XC-A30 was the tested method. Wilcoxon signed rank test showed no difference between two methods (p= 0.439). There was a very strong correlation and excellent agreement showed by Spearman coefficient correlation r= 0.989 and kappa coefficient 0.942. Conclusion: This study showed no difference between Caretium XC-A30 and Westergren as the reference method for ESR measurement, with very strong correlation and excellent agreement. Thus, Caretium XC-A30 has a potential use in performing ESR measurement with high throughput and lesser time than manual Westergren method.
机译:背景:红细胞沉降率(ESR)测试是评估炎症,感染,自身免疫或恶性肿瘤急性期的最广泛执行的实验室测试之一。 Westergren介绍了用于ESR测量的参考方法。该方法通常是简单,廉价的,但是很耗时。引入了自动仪器以减少测量时间。 Caretium XC-A30是自动测量ESR的仪器;未经先前研究,它已在Sanglah总医院登巴萨使用。这项研究的目的是评估Caretium XC-A30与参考方法的比较,相关性和一致性。方法:2018年8月在登巴萨Sanglah总医院采集了35份连续样本。用Caretium XC-A30和手册Westergren检查样品的ESR。比较了两种方法的比较,相关性和一致性。结果:Westergren方法为参考方法,Caretium XC-A30为测试方法。 Wilcoxon符号秩检验显示两种方法之间没有差异(p = 0.439)。 Spearman系数相关性r = 0.989和kappa系数0.942显示出非常强的相关性和极好的一致性。结论:本研究表明,Caretium XC-A30与Westergren作为ESR测量的参考方法之间没有差异,具有很强的相关性和良好的一致性。因此,与手动Westergren方法相比,Caretium XC-A30具有高通量和更少时间的ESR测量潜力。

著录项

相似文献

  • 外文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号