首页> 外文期刊>Biomedical Optics Express >Effect of signal intensity and camera quantization on laser speckle contrast analysis
【24h】

Effect of signal intensity and camera quantization on laser speckle contrast analysis

机译:信号强度和相机量化对激光散斑对比度分析的影响

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Laser speckle contrast analysis (LASCA) is limited to being a qualitative method for the measurement of blood flow and tissue perfusion as it is sensitive to the measurement configuration. The signal intensity is one of the parameters that can affect the contrast values due to the quantization of the signals by the camera and analog-to-digital converter (ADC). In this paper we deduce the theoretical relationship between signal intensity and contrast values based on the probability density function (PDF) of the speckle pattern and simplify it to a rational function. A simple method to correct this contrast error is suggested. The experimental results demonstrate that this relationship can effectively compensate the bias in contrast values induced by the quantized signal intensity and correct for bias induced by signal intensity variations across the field of view.
机译:激光散斑对比度分析(LASCA)限于测量血流和组织灌注的定性方法,因为它对测量配置敏感。信号强度是由于相机和模数转换器(ADC)对信号进行量化而可能影响对比度值的参数之一。在本文中,我们基于斑点图案的概率密度函数(PDF)推导了信号强度和对比度值之间的理论关系,并将其简化为有理函数。建议了一种纠正此对比度误差的简单方法。实验结果表明,这种关系可以有效补偿量化信号强度引起的对比度值偏差,并校正整个视场中信号强度变化引起的偏差。