摘要

The remainder of H2O2, which is usually used in advanced oxidation processes (AOPs) treatment of wastewaters, reacts with the oxidant (e.g. KMnO4) utilized in chemical oxygen demand (COD) analysis, resulting in the overestimation in COD analysis. The residual H2O2 was decomposed successfully using Na2CO3 as catalyst under a heated condition (95 degrees C for 2h) and consequently the H2O2-derived COD, typically after the AOPs treatment of oxalic acid solution using peroxone (H2O2-O-3 mixture), could be brought to below the detection limit. However, the overdose of acid (H2SO4) added for neutralizing the Na2CO3-containing sample solution after the decomposition of residual H2O2 was found to cause new interference in COD analysis. It is considered that in strong acid solutions the decomposition of KMnO4 may be accelerated during COD analysis, leading to its overestimation. The results suggest that a suitable decomposition of residual H2O2 (e.g. after AOPs treatment) as well as a suitable control of acidity (to neutrality) of the sample solution subjected to COD analysis is essentially necessary for the correct COD estimation.
机译:通常用于高级氧化过程(AOP)处理废水的H2O2的其余部分与化学需氧量(COD)分析中使用的氧化剂(例如KMNO4)反应,导致COD分析中的高估。在加热条件下使用Na 2 CO 3作为催化剂成功地分解残留的H 2 O 2,并且因此使用过氧化物(H2O2-3混合物)的AOP处理草酸溶液后,可以被带到检测限下。然而,在发现残留H2O2分解后,添加用于中和用于中和含Na 2 CO 3的样品溶液的酸(H 2 SO 4)造成新的干扰在COD分析中。认为,在强酸溶液中,KMnO4的分解可以在COD分析期间加速,导致其高估。结果表明,残留H 2 O 2的合适分解(例如,AOP处理之后)以及对进行COD分析的样品溶液的酸度(至中性)的合适控制基本上是正确的COD估计所必需的。

著录项

  • 来源
    《Desalination and water treatment》 |2016年第56期|a1927138-27143|共7页
  • 作者单位

    Tokyo Inst Technol Interdisciplinary Grad Sch Sci & Engn Dept Elect Chem Midori Ku 4259-G1-5 Nagatsuta Yokohama Kanagawa 2268502 Japan;

    Tokyo Inst Technol Interdisciplinary Grad Sch Sci & Engn Dept Elect Chem Midori Ku 4259-G1-5 Nagatsuta Yokohama Kanagawa 2268502 Japan;

    Tokyo Inst Technol Interdisciplinary Grad Sch Sci & Engn Dept Elect Chem Midori Ku 4259-G1-5 Nagatsuta Yokohama Kanagawa 2268502 Japan;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Na2CO3; Oxalic acid; COD; H2O2 interference; Water treatment;

    机译:Na2CO3;草酸;COD;H2O2干扰;水处理;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号