The influence of different magnitudes and methods of applying preload on fusion and disc replacement constructs in the lumbar spine: a finite element analysis

机译:不同大小和施加预紧力的方法对腰椎融合和椎间盘置换结构的影响:有限元分析

获取原文
获取原文并翻译 | 示例
           

摘要

In a finite element (FE) analysis of the lumbar spine, different preload application methods that are used in biomechanical studies may yield diverging results. To investigate how the biomechanical behaviour of a spinal implant is affected by the method of applying the preload, hybrid-controlled FE analysis was used to evaluate the biomechanical behaviour of the lumbar spine under different preload application methods. The FE models of anterior lumbar interbody fusion (ALIF) and artificial disc replacement (ADR) were tested under three different loading conditions: a 150N pressure preload (PP) and 150 and 400 N follower loads (FLs). This study analysed the resulting range of motion (ROM), facet contact force (FCF), inlay contact pressure (ICP) and stress distribution of adjacent discs. The FE results indicated that the ROM of both surgical constructs was related to the preload application method and magnitude; differences in the ROM were within 7% for the ALIF model and 32% for the ADR model. Following the application of the FL and after increasing the FL magnitude, the FCF of the ADR model gradually increased, reaching 45% at the implanted level in torsion. The maximum ICP gradually decreased by 34.1% in torsion and 28.4% in lateral bending. This study concluded that the preload magnitude and application method affect the biomechanical behaviour of the lumbar spine. For the ADR, remarkable alteration was observed while increasing the FL magnitude, particularly in the ROM, FCF and ICP. However, for the ALIF, PP and FL methods had no remarkable alteration in terms of ROM and adjacent disc stress.
机译:在腰椎的有限元分析中,生物力学研究中使用的不同预紧力施加方法可能会产生不同的结果。为了研究预紧力施加方法对脊柱植入物的生物力学行为的影响,采用混合控制有限元分析评估了不同预紧力施加方法下腰椎的生物力学行为。前腰椎椎间融合术(ALIF)和人工椎间盘置换术(ADR)的有限元模型在三种不同的加载条件下进行了测试:150N压力预加载(PP)和150和400 N随动加载(FL)。这项研究分析了所产生的运动范围(ROM),小平面接触力(FCF),嵌体接触压力(ICP)和相邻光盘的应力分布。有限元分析结果表明,两种手术结构的ROM都与预加载方法和大小有关。对于ALIF模型,ROM中的差异在7%之内,对于ADR模型,其ROM中的差异在32%之内。在应用FL后,在增加FL幅度之后,ADR模型的FCF逐渐增加,在植入的扭转水平达到45%。最高ICP扭转时逐渐减小34.1%,横向弯曲时逐渐减小28.4%。这项研究得出的结论是,预紧力大小和施加方法会影响腰椎的生物力学行为。对于ADR,在增加FL幅度的同时观察到了明显的变化,特别是在ROM,FCF和ICP中。但是,对于ALIF,PP和FL方法在ROM和相邻光盘应力方面没有明显改变。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号