Endovascular stent configuration affects intraluminal flow dynamics and in vitro endothelialization

机译:血管内支架配置影响腔内流动动力学和体外内皮化

获取原文
获取原文并翻译 | 示例
       

摘要

Neointimal hyperplasia influenced by intravascular hemodynamics is considered partly responsible for restenosis after endovascular stenting. To evaluate the effect of stent configuration on fluid flow behavior, we visualized flow near stents, and measured the proliferation of cultured endothelial cells (ECs). A single-coil stent (coil pitch; CP = 2.5, 5, or 10 mm) was inserted into a glass tube and perfused at 30-90 ml/min, while the flow pattern was determined by particle imaging velocimetry. The reduction of the flow velocity near the wall was correlated with the decrease in the coil interval of the stent. In perfusion cultures with stents, the proliferation of ECs was influenced by the local flow velocity distribution. When a stent with a CP value of 10 mm was used, the doubling time of ECs was 30.7 h, while the doubling time was 38.5 h when the CP was 5 mm. The doubling time of ECs was shorter at sites upstream of the stent wire where the velocity was higher than downstream of the wire. In conclusion, a single-coil stent can be used to modify hemodynamic factors, suggesting that improved stent design may facilitate rapid endothelialization after stent implantation.
机译:受血管内血流动力学影响的新内膜增生被认为部分归因于血管内支架置入术后的再狭窄。为了评估支架配置对流体流动行为的影响,我们将支架附近的流动可视化,并测量了培养的内皮细胞(EC)的增殖。将单线圈支架(线圈间距; CP = 2.5、5或10 mm)插入玻璃管中,并以30-90 ml / min的速度灌注,同时通过颗粒成像测速仪确定流动模式。壁附近流速的降低与支架线圈间隔的减少有关。在带支架的灌注培养中,ECs的增殖受局部流速分布的影响。当使用CP值为10 mm的支架时,EC的倍增时间为30.7 h,而CP为5 mm时的倍增时间为38.5 h。 ECs的倍增时间在支架线上游的位置处的速度要短于在支架线下游的位置。总之,可以使用单线圈支架来改变血液动力学因素,这表明改进的支架设计可以促进支架植入后的快速内皮化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号