首页> 外文期刊>IEEE Transactions on Biomedical Engineering >Removal of CPR Artifacts From the Ventricular Fibrillation ECG by Adaptive Regression on Lagged Reference Signals
【24h】

Removal of CPR Artifacts From the Ventricular Fibrillation ECG by Adaptive Regression on Lagged Reference Signals

机译:通过对滞后参考信号进行自适应回归从心室颤动心电图去除CPR伪像

获取原文
获取原文并翻译 | 示例

摘要

Background and objective: Removing cardiopulmonary resuscitation (CPR)-related artifacts from human ventricular fibrillation (VF) electrocardiogram (ECG) signals provides the possibility to continuously detect rhythm changes and estimate the probability of defibrillation success. This could reduce “hands-off” analysis times which diminish the cardiac perfusion and deteriorate the chance for successful defibrillations. Methods and results: Our approach consists in estimating the CPR part of a corrupted signal by adaptive regression on lagged copies of a reference signal which correlate with the CPR artifact signal. The algorithm is based on a state-space model and the corresponding Kalman recursions. It allows for stochastically changing regression coefficients. The residuals of the Kalman estimation can be identified with the CPR-filtered ECG signal. In comparison with ordinary least-squares regression, the proposed algorithm shows, for low signal-to-noise ratio (SNR) corrupted signals, better SNR improvements and yields better estimates of the mean frequency and mean amplitude of the true VF ECG signal. Conclusions: The preliminary results from a small pool of human VF and animal asystole CPR data are slightly better than the results of comparable previous studies which, however, not only used different algorithms but also different data pools. The algorithm carries the possibility of further optimization.
机译:背景与目的:从人心室纤颤(VF)心电图(ECG)信号中去除与心肺复苏(CPR)相关的伪像,为不断检测心律变化并评估除颤成功的可能性提供了可能性。这可以减少“手动”分析时间,从而减少心脏灌注并降低成功除颤的机会。方法和结果:我们的方法包括通过对与CPR伪影信号相关的参考信号的滞后副本进行自适应回归来估计损坏信号的CPR部分。该算法基于状态空间模型和相应的Kalman递归。它允许随机改变回归系数。卡尔曼估计的残差可以通过CPR滤波的ECG信号进行识别。与普通的最小二乘回归相比,该算法表明,对于低信噪比(SNR)损坏的信号,SNR改善得更好,并且可以对真实VF ECG信号的平均频率和平均幅度进行更好的估计。结论:一小部分人VF和动物心搏过CPR数据的初步结果比以前的可比研究更好,后者不仅使用了不同的算法,而且还使用了不同的数据库。该算法具有进一步优化的可能性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号