A novel extrapolation method using OSL detectors for very small field output factor measurement for stereotactic radiosurgery

机译:一种新的外推方法,使用OSL探测器对非常小的田间输出因子测量对刻膜放射前的

获取原文
获取原文并翻译 | 示例
           

摘要

Appropriate methods for the determination of very small X-ray beam output factors are essential to ensure correct clinical outcomes for stereotactic radiosurgery. To date, substantial work has been performed in identifying and quantifying suitable dosimeters for relative output factor (ROF) measurements including recent IAEA published recommendations. In this work, we provide a novel method using optically stimulated luminescent dosimeters (OSLDs) with different effective sizes of the readout area to determine ROFs. This involves applying an extrapolation technique to assess ROFs for 6MV SRS X-ray beams with field diameters ranging from 4 to 30 mm as defined by the Brainlab SRS cones. By combining the use of multiple sized OSLDs and water droplets to remove air gaps located around the OSLD detectors, both volume averaging and density variation effects were minimised to estimate ROFs for an extrapolated zero volume detector. The measured results showed that for a 4 mm diameter cone, the ROF was 0.660 +/- 0.032 (2SD) as compared to 0.661 +/- 0.01 and 0.651 +/- 0.018 for the PTW 600019 microDiamond detector and Gafchromic EBT3 film respectively. Whilst the uncertainties were larger than conventional detectors, the technique shows promise and improvements in accuracy may be obtained by higher quality manufacturing techniques. Based on these results, using OSLDs with different effective sizes of readout area and an extrapolation technique shows promise for use as an independent verification tool for very small X-ray field ROFs in the clinical department.
机译:确定非常小的X射线束输出因子的适当方法对于确保立体定向放射牢房的正确临床结果是必不可少的。迄今为止,在识别和量化相对输出因子(ROF)测量的合适剂量计中,已经进行了实质性的作品,包括最近的原子能机构出版的建议。在这项工作中,我们提供一种新的方法,使用光学刺激的发光剂量计(OSLD)具有不同的有效尺寸的读出区域来确定ROF。这涉及应用外推技术,以评估6mV SRS X射线束的ROF,其场直径范围为4至30mm,如脑锥体所定义的。通过组合多个尺寸的OSLD和水滴去除位于OSLD探测器周围的空气间隙,最小化体积平均和密度变化效应,以估计外推零体积检测器的ROF。测量结果表明,对于4mm直径的锥体,ROF分别为0.661 +/- 0.032(2SD),分别为0.661 +/- 0.01和0.651 +/- 0.018和0.651 +/- 0.018,分别为0.661 +/- 0.018和GafChromic EBT3膜。虽然不确定因素大于传统探测器,但是该技术可以通过更高质量的制造技术来获得准确性的承诺和改进。基于这些结果,使用具有不同有效读数区域的OSLD和外推技术,显示了用作临床部中非常小的X射线场rofs的独立验证工具。

著录项

  • 来源
  • 作者单位

    Univ Sydney Sch Phys Inst Med Phys Sydney NSW 2006 Australia;

    Chris OBrien Lifehouse Dept Radiat Oncol 119-143 Missenden Rd Camperdown NSW 2050 Australia|Queensland Univ Technol Sch Chem Phys & Mech Engn Level 4 O Block Brisbane Qld 4001 Australia;

    Univ Sydney Sch Phys Inst Med Phys Sydney NSW 2006 Australia;

    Univ Sydney Sch Phys Inst Med Phys Sydney NSW 2006 Australia;

    Univ Sydney Sch Phys Inst Med Phys Sydney NSW 2006 Australia|Chris OBrien Lifehouse Dept Radiat Oncol 119-143 Missenden Rd Camperdown NSW 2050 Australia;

    Univ Sydney Sch Phys Inst Med Phys Sydney NSW 2006 Australia|Chris OBrien Lifehouse Dept Radiat Oncol 119-143 Missenden Rd Camperdown NSW 2050 Australia;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Small X-ray field dosimetry; OSLD; Extrapolation technique;

    机译:小X射线场剂量测定法;OSLD;外推技术;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号