摘要

The reference air kerma rate from ~(192)Ir High Dose Rate (HDR) brachytherapy sources can be measured using a suitably calibrated Farmer chamber and an appropriate in-air calibration jig. When a primary standard for ~(192)Ir gamma rays is available, a calibration coefficient for the chamber and jig combination can be determined directly. In Australia, due to the absence of such a standard, the chamber must be calibrated by interpolation of the response in ~(60)Co and in a kilovoltage x-ray beam. Corrections for the effect of the jig, scatter and beam non-uniformity must then be measured or calculated before the reference air kerma rate can be determined. We compare the air-kerma calibration coefficient of a PTW 30010 PMMA/A1 Farmer chamber (referred to as Fanner chamber throughout this report) obtained from the ~(192)Ir primary standard at the National Physical Laboratory in the UK with the corresponding coefficient obtained by interpolating Australian calibrations using ~(60)Co and 250 kV x-rays and determining suitable correction factors. The resulting chamber/jig calibration coefficients differ by 0.2% which is well within the combined standard uncertainties of 1.2% and 0.6% reported by ARPANSA and NPL respectively.
机译:来自〜(192)Ir高剂量率(HDR)近距离放射治疗源的参考空气比释动能比率可以使用适当校准的Farmer箱和适当的空气校准夹具进行测量。当可获得〜(192)Ir伽马射线的主要标准时,可以直接确定腔室和夹具组合的校准系数。在澳大利亚,由于缺乏这种标准,必须通过内插〜(60)Co和千伏X射线束中的响应来校准反应室。在确定参考空气比释动能比之前,必须先测量或计算出夹具,散射和光束不均匀性的影响的校正值。我们比较了从英国国家物理实验室的〜(192)Ir一级标准获得的PTW 30010 PMMA / A1农夫舱(在本报告中称为Fanner舱)的空气比释动能校准系数与相应的系数通过使用〜(60)Co和250 kV X射线对澳大利亚校准进行插值并确定合适的校正因子。所得的腔/夹具校准系数相差0.2%,这恰好在ARPANSA和NPL分别报告的1.2%和0.6%的组合标准不确定度之内。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号