A retrospective evaluation of radiation dose associated with low dose FDG protocols in whole-body PET/CT

机译:对全身PET / CT中低剂量FDG方案相关的放射剂量的回顾性评估

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The objective of the study is to retrospectively measure patient radiation dose resulting from whole body X-ray CT and FDG PET studies using a low-dose protocol performed on the Siemens Biograph mCT scanner. A total of 483 patient studies were reviewed. For each, the CT dose-length product was used to estimate radiation dose to the patient as a result of the whole body X-ray CT component of the PET/CT study. The net injected FDG dose was used to calculate the whole body effective dose based on ICRP recommendations. Dose calculations were also modified to take into account individual patient weight. The total effective dose received by each patient was taken as the sum of the PET and CT components of the study. The mean effective dose from the CT component of the diagnostic study was found to be 8.2 mSv (3.4-23.4 mSv), for a CT protocol of 120 kV_p and effective tube current-time product of 80 mAs with automatic exposure control. For an average injected FDG activity of 304 MBq the mean PET effective dose was found to be 6.3 mSv when using the ICRP standard models, or 6.0 mSv when scaling effective dose to individual patient weight or patient blood volume. The average total effective dose across the entire patient cohort for a combined PET/CT study was found to be ~14.5 mSv (9.6-29.8 mSv). Low-dose protocols for whole-body PET/CT scanning result in an effective radiation dose to the patient of approximately 14.5 mSv. Additional reductions through the use of iterative CT reconstruction and optimized low-dose FDG protocols could see total effective doses for whole-body PET/CT fall to below 10 mSv.
机译:该研究的目的是使用在西门子Biograph mCT扫描仪上执行的低剂量方案,回顾性地测量由全身X射线CT和FDG PET研究得出的患者辐射剂量。总共审查了483例患者研究。对于每种剂量,由于PET / CT研究的全身X射线CT成分,CT剂量长度乘积用于估计对患者的辐射剂量。净注射的FDG剂量用于根据ICRP建议计算全身有效剂量。剂量计算也进行了修改,以考虑个体患者的体重。每位患者接受的总有效剂量被视为研究中PET和CT成分的总和。对于120 kV_p的CT协议和80 mAs的有效管电流-时间乘积(具有自动曝光控制),诊断研究的CT组件得出的平均有效剂量为8.2 mSv(3.4-23.4 mSv)。对于平均注射的FDG活性为304 MBq,使用ICRP标准模型时,平均PET有效剂量为6.3 mSv,而根据个体患者体重或患者血容量换算有效剂量时,平均PET有效剂量为6.0 mSv。整个PET / CT研究的整个患者队列的平均总有效剂量为〜14.5 mSv(9.6-29.8 mSv)。用于全身PET / CT扫描的小剂量方案可为患者带来约14.5 mSv的有效辐射剂量。通过使用迭代CT重建和优化的低剂量FDG方案进一步减少,可以使全身PET / CT的总有效剂量降至10 mSv以下。

著录项

  • 来源
  • 作者单位

    Institute of Medical Physics, University of Sydney,Sydney, Australia;

    Department of Nuclear Medicine, Royal North Shore Hospital,St Leonards 2065, Sydney, NSW, Australia,Faculty of Health Sciences, University of Sydney,Sydney, Australia;

    Department of Nuclear Medicine, Royal North Shore Hospital,St Leonards 2065, Sydney, NSW, Australia,Faculty of Health Sciences, University of Sydney,Sydney, Australia;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    FDG; dosimetry; PET; CT;

    机译:FDG;剂量学宠物;电脑断层扫描;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号