In detecting monoenergetic γ rays, the instrumental line of the scintillation spectrometer consists of a photopeak cor- responding to the photons absorbed in the crystal and a wide distribution due to photons which were scattered in the detector and left the detector losing only a part of their energy [1]. Ordinarily, the energy of the primary photons is judged according to the position of a photopeak.
机译:在检测单能γ射线时,闪烁光谱仪的仪器线由与晶体中吸收的光子相对应的光峰和由于光子分散在检测器中而使检测器仅损失一部分光子而引起的宽分布所组成。能量[1]。通常,根据光峰的位置来判断一次光子的能量。