首页> 外文期刊>Atomic Energy >UPDATED STATE PRIMARY SPECIAL STANDARD OF UNITS OF NEUTRON FLUX DENSITY AND NEUTRON FLUENCE FOR NUCLEAR-PHYSICS FACILITIES
【24h】

UPDATED STATE PRIMARY SPECIAL STANDARD OF UNITS OF NEUTRON FLUX DENSITY AND NEUTRON FLUENCE FOR NUCLEAR-PHYSICS FACILITIES

机译:核物理设施的中子通量密度和中子通量单位更新的国家主要特殊标准

获取原文
获取原文并翻译 | 示例
           

摘要

The design of updated sources of neutrons from the composition of the State Primary Special Standard of units of neutron flux density and neutron fluence for nuclear-physics facilities, GET 51-17, whose upper bound of the reproduction range is raised several-fold as a result of the use of a better neutron generator - NG-10M, is described. The principles and methods of reproducing the neutron flux density and neutron fluence units on a reference standard and evaluating accuracy are described. The metrological characteristics of the updated reference standard, which were obtained in the course of its study and during primary certification, are presented and the sources of error of the reference standard are described in detail.
机译:根据《国家主要特殊标准》的组成,设计用于核物理设施的中子通量密度和中子注量单位的更新中子源,GET 51-17,其再生范围的上限提高了几倍。描述了使用更好的中子发生器NG-10M的结果。描述了在参考标准上再现中子通量密度和中子注量单位并评估准确性的原理和方法。介绍了在研究过程中和初次认证过程中获得的最新参考标准的计量特性,并详细描述了参考标准的误差来源。