首页> 外文期刊>Applied Superconductivity, IEEE Transactions on >An 800-MHz all-REBCO Insert for the 1.3-GHz LTS/HTS NMR Magnet Program—A Progress Report
【24h】

An 800-MHz all-REBCO Insert for the 1.3-GHz LTS/HTS NMR Magnet Program—A Progress Report

机译:用于1.3 GHz LTS / HTS NMR磁体程序的800 MHz全REBCO插入件-进展报告

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A critical component of the 1.3-GHz nuclear magnetic resonance magnet (1.3 G) program, currently ongoing at the Francis Bitter Magnet Laboratory, Plasma Science and Fusion Center, Massachusetts Institute of Technology, and now approaching its final stage, is the all high-temperature superconductor 800-MHz insert (H800). The insert consists of three nested double-pancake (DP) coils fabricated with 6-mm-wide REBCO conductor. Coil 1, the innermost coil of H800, has already been fabricated and tested at 77 and 4.2 K. In addition, one third of the DPs for Coil 2 have been wound and each DP individually fully tested. Work described here includes details of Coil 1 fabrication: DP winding, DP testing, assembling, joint performance, overbanding, and coil testing; winding details of DPs for Coil 2 and their testing are also included.
机译:目前,1.3 GHz核磁共振磁体(1.3 G)计划的关键组成部分是全高水平的研究,该计划目前正在麻省理工学院等离子科学与聚变中心的弗朗西斯·波特磁体实验室进行,目前已接近尾声。温度超导体800 MHz插件(H800)。插入件由三个嵌套的双薄饼(DP)线圈组成,这些线圈由6毫米宽的REBCO导体制成。 H800的最里面的线圈1已经在77和4.2 K的温度下进行了制造和测试。另外,线圈2的DP的三分之一已经缠绕,并且每个DP都经过了充分测试。这里描述的工作包括Coil 1制造的细节:DP绕组,DP测试,组装,接头性能,超频和线圈测试;还包括线圈2的DP的绕组详细信息及其测试。