Ammonia trace measurements at ppb level based on near-IR photoacoustic spectroscopy

机译:基于近红外光声光谱法的ppb级氨微量测量

获取原文
获取原文并翻译 | 示例
           

摘要

A photoacoustic sensor using a laser diode emitting near 1532 nm in combination with an erbium-doped fibre amplifier has been developed for ammonia trace gas analysis at atmospheric pressure. NH_3 concentration measurements down to 6 ppb and a noise-equivalent detection limit below 3 ppb in dry air are demonstrated. Two wavelength-modulation schemes with 1f and 2f detection using a lock-in amplifier were investigated and compared to maximise the signal-to-noise ratio. A quantitative analysis of CO_2 and H_2O interference with NH_3 is presented. Typical concentrations present in ambient air of 400 ppm CO_2 and 1.15% H_2O (50% relative humidity at 20℃) result in a NH_3 equivalent concentration of 36 ppb and 100 ppb, respectively.
机译:已经开发出一种使用发射接近1532 nm的激光二极管和掺fiber光纤放大器的光声传感器,用于在大气压力下分析氨气。演示了干燥空气中NH_3浓度测量值低至6 ppb和等效噪声检测极限低于3 ppb的情况。研究并比较了使用锁定放大器检测1f和2f的两种波长调制方案,以最大化信噪比。给出了CO_2和H_2O对NH_3干扰的定量分析。在环境空气中存在的典型浓度为400 ppm CO_2和1.15%H_2O(在20℃时相对湿度为50%)会导致NH_3当量浓度分别为36 ppb和100 ppb。