首页> 外文期刊>Applied physics >Multi-harmonic detection in wavelength modulation spectroscopy systems
【24h】

Multi-harmonic detection in wavelength modulation spectroscopy systems

机译:波长调制光谱系统中的多谐波检测

获取原文
获取原文并翻译 | 示例
           

摘要

Two multi-harmonic detection methods for wavelength modulation spectroscopy (WMS) systems are presented and compared. The two possibilities discussed in this paper are: simultaneous curve fitting of multiple harmonic spectra, and reconstruction of the transmission from harmonic coefficients. The optimum number of harmonics is four and 25 harmonics, respectively. Compared with standard single-harmonic curve fitting, the methods give about a factor of 3 better performance than standard second-harmonic curve fitting. Concluding, multi-harmonic detection is better than single-harmonic detection and should be used if the system bandwidth is high enough to allow for proper detection of the higher harmonics.
机译:提出并比较了两种用于波长调制光谱(WMS)系统的多谐波检测方法。本文讨论的两种可能性是:多个谐波频谱的同时曲线拟合,以及从谐波系数重构传输。最佳谐波数分别为4和25。与标准的单谐波曲线拟合相比,该方法的性能要比标准的第二谐波曲线拟合好3倍。最后,多谐波检测优于单谐波检测,如果系统带宽足够高,可以正确检测较高的谐波,则应使用多谐波检测。

著录项

  • 来源
    《Applied physics》 |2013年第2期|177-185|共9页
  • 作者单位

    Walter Schottky Institute, Technical University of Munich,Am Coulombwall 3, 85748 Garching, Germany,Siemens Corporate Research & Technologies, Siemens AG,Otto-Hahn-Ring 6, 80200 Munich, Germany;

    Walter Schottky Institute, Technical University of Munich,Am Coulombwall 3, 85748 Garching, Germany;

    Siemens Corporate Research & Technologies, Siemens AG,Otto-Hahn-Ring 6, 80200 Munich, Germany;

    Walter Schottky Institute, Technical University of Munich,Am Coulombwall 3, 85748 Garching, Germany;

  • 收录信息 美国《科学引文索引》(SCI);美国《工程索引》(EI);美国《生物学医学文摘》(MEDLINE);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号