This study investigates the performance of a new statistically driven acute ischemia detection algorithm that can process data from two bipolar cutaneous or subcutaneous leads. During a start-up phase, the algorithm processes electrocardiogram signals to determine a normal range of ST-segment deviation as a function of heart rate. The algorithm then generates upper and lower ST-deviation thresholds based on the dispersion of the baseline ST-deviation data. After the start-up phase, persistent ST-deviation that is beyond either the upper or lower thresholds results in detection of acute ischemia. To test the algorithm, we performed long-term (10 day) Holter monitoring in a control group of 14 subjects. We also performed Holter monitoring during balloon angioplasty, and for 2 days after surgery, in 30 subjects who underwent elective percutaneous coronary interventions (“PCI”). We determined the percentage of balloon inflations the algorithm detected without producing false positive detections within the control group 10-day daily life data. The algorithm detected 17/17 LAD occlusions, 7/8 LCX occlusions, and 8/9 RCA occlusions. Our results suggest that automatically generated, subject-specific, heart-rate dependent ST-deviation thresholds can detect PCI induced myocardial ischemia without resulting in false positive detections in a small control group.
机译:这项研究调查了一种新的统计驱动的急性缺血检测算法的性能,该算法可以处理来自两个双极皮肤或皮下导线的数据。在启动阶段,该算法处理心电图信号以确定ST段偏差的正常范围作为心率的函数。然后,该算法根据基线ST偏差数据的离散度生成上下ST偏差阈值。在启动阶段之后,超过上限或下限的持续性ST偏离会导致急性缺血的检测。为了测试算法,我们在14名受试者的对照组中进行了长期(10天)动态心电图监测。我们还对30名接受了择期经皮冠状动脉介入治疗(“ PCI”)的受试者在球囊血管成形术期间和术后2天进行了动态心电图监测。我们确定了在对照组的10天日常生活数据中,算法检测到的气球充气百分比,未产生假阳性检测结果。该算法检测到17/17 LAD遮挡,7/8 LCX遮挡和8/9 RCA遮挡。我们的结果表明,自动生成的,受测者特定的,依赖于心率的ST偏差阈值可以检测PCI诱发的心肌缺血,而不会在一个小的对照组中导致假阳性检测。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号