首页> 外文期刊>Age and Ageing >Recurrent strokes caused by a malpositioned pacemaker lead
【24h】

Recurrent strokes caused by a malpositioned pacemaker lead

机译:起搏器导线位置错误导致的中风复发

获取原文
获取原文并翻译 | 示例
       

摘要

This case report illustrates the case of a patient who developed recurrent strokes after a pace maker lead was inserted into his left ventricle. It was removed successfully by the cardiothoracic surgeons but he remained very dependant functionally. This case highlights the importance of always reviewing the electrocardiogram and chest radiograph after the insertion of a pacemaker as late diagnosis of this complication can leave the patient with significant morbidity.
机译:该病例报告说明了一名患者,该患者在将起搏器导线插入其左心室后发生中风复发。心胸外科医师成功地将其切除,但他在功能上仍然非常依赖。该病例突出显示了在插入起搏器后务必复查心电图和胸部X光片的重要性,因为这种并发症的晚期诊断可能会使患者罹患严重疾病。

著录项

  • 来源
    《Age and Ageing》 |2012年第3期|p.420-421|共2页
  • 作者

    Chakravarthi Rajkumar;

  • 作者单位
  • 收录信息 美国《科学引文索引》(SCI);美国《化学文摘》(CA);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

  • 入库时间 2022-08-18 01:10:25

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号