Accuracy and Precision of a Veterinary Neuronavigation System for Radiation Oncology Positioning

机译:兽医神经导航系统在放射肿瘤学中的定位精度

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Conformal radiation treatment plans such as IMRT and other radiosurgery techniques require very precise patient positioning, typically within a millimeter of error for best results. CT cone beam, real-time navigation, and infrared position sensors are potential options for success but rarely present in veterinary radiation centers. A neuronavigation system (Brainsight Vet, Rogue Research) was tested 22 times on a skull for positioning accuracy and precision analysis. The first 6 manipulations allowed the authors to become familiar with the system but were still included in the analyses. Overall, the targeting mean error in 3D was 1.437 mm with SD 1.242 mm. This system could be used for positioning for radiation therapy or radiosurgery.
机译:适形放射治疗计划(例如IMRT和其他放射外科技术)需要非常精确的患者定位,通常要在误差范围内达到一毫米才能获得最佳效果。 CT锥形束,实时导航和红外位置传感器是成功的潜在选择,但在兽医辐射中心很少出现。在颅骨上对神经导航系统(Brainsight Vet,Rogue Research)进行了22次测试,以进行定位精度和精度分析。前6种操作允许作者熟悉该系统,但仍包含在分析中。总体而言,使用SD 1.242mm时,3D的目标平均误差为1.437mm。该系统可以用于放射治疗或放射外科手术的定位。