Automated Monitoring of the Establishment of the Adsorption Equilibrium: Adsorption of Polyethylene from 124-Trichlorobenzene onto a Zeolite at Temperature 140°C

机译:自动化监测吸附平衡的建立:温度为140°C聚乙烯从124-三氯苯吸附到沸石上

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The automated procedure for the monitoring of the adsorption process in the solute-sorbent-solvent system has been elaborated. It uses commercially available instrument CRYSTAF model 200. The application of CRYSTAF enabled monitoring of adsorption of linear polyethylene with weight average molar masses of 2, 14, and 53 kg/mol from 1,2,4-trichlorobenzene onto zeolite SH-300 at temperature as high as 140°C. It is the authors' understanding that this is the first demonstration of an adsorption isotherms for polyethylene. The measurement with the CRYSTAF instrument reduces manual manipulations with dangerous solvents at high temperature and enables automated long-time monitoring of the concentration of the solute in an adsorption system.
机译:已经制定了监测溶质-吸附剂-溶剂系统中吸附过程的自动化程序。它使用市售的200型仪器CRYSTAF。CRYSTAF的应用使得能够监测重均摩尔质量分别为2、14和53 kg / mol的线性聚乙烯从1,2,4-三氯苯在温度为1,2,4-三氯苯上的吸附高达140°C。作者的理解是,这是聚乙烯吸附等温线的第一个证明。使用CRYSTAF仪器进行的测量减少了危险溶剂在高温下的手动操作,并能够自动长期监控吸附系统中溶质的浓度。