A novel approach to non-segmented flow analysis: Part 3. Nitrate nitrite and ammonium in waters

机译:一种非分段流量分析的新颖方法:第3部分。水中的硝酸盐亚硝酸盐和铵盐

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A high-performance continuous-flow analyser is used for the analysis of nitrate, nitrite and ammonium ion in potable waters. The results demonstrate that (1) the analyser allows the sequential determination of a number of analytes without requiring modification to the manifold; (2) the use of programmed slicing of the reaction mixture allows a wide range of analyte concentrations to be handled; and (3) that the sensitivity achieved compares favourably with the best available from conventional flow-injection analysis. The limits of detection were found to be 5 ppb for NH4 +, 30 ppb for NO3 -, and 4 ppb for NO2 -.
机译:高性能连续流分析仪用于分析饮用水中的硝酸盐,亚硝酸盐和铵离子。结果表明:(1)分析仪无需改变歧管即可连续测定许多分析物; (2)使用反应混合物的程序切片可以处理各种浓度的分析物; (3)获得的灵敏度与常规流动注射分析所能获得的最佳灵敏度相媲美。 NH4 + 的检出限为5 ppb,NO3 -的检出限为30 ppb,NO2 -的检出限为4 ppb。