Two flow methods, based, respectively, on flow-injection analysis (FIA) and on multicommutated flow analysis (MCFA), were compared with regard to their use for the determination of total selenium in infant formulas by hydride-generation atomic absorption spectrometry. The method based on multicommutation provided lower detection and quantification limits (0.08 and 0.27 μg L−1 compared to 0.59 and 1.95 μ L−1, resp.), higher sampling frequency (160 versus. 70 samples per hour), and reduced reagent consumption. Linearity, precision, and accuracy were similar for the two methods compared. It was concluded that, while both methods proved to be appropriate for the purpose, the MCFA-based method exhibited a better performance.
机译:比较了分别基于流动注射分析(FIA)和多换向流动分析(MCFA)的两种流动方法,用于氢化物发生原子吸收光谱法测定婴儿配方食品中的总硒。基于多重换向的方法检测和定量限较低(分别为0.08和0.27μgL -1 分别为0.59和1.95μgL -1 ),较高的采样频率(每小时160个样品与70个样品),并减少了试剂消耗。两种方法的线性,精度和准确性相似。结论是,尽管两种方法均证明适用于该目的,但基于MCFA的方法表现出更好的性能。