摘要

Endovascular particle embolization is a common procedure with a relatively safe profile. We report here four cases in which cranial nerves, skin and mucosal ischemic complications occurred with the use of hydrogel microspheres (250–500 µm in size). Given the compressibility and higher penetration potential of microsphere particles compared with polyvinyl alcohol particles of similar size, we suggest oversizing hydrogel microsphere particles for head and neck embolizations.
机译:血管内颗粒栓塞是具有相对安全轮廓的常见程序。我们在这里报告了四例使用水凝胶微球体(大小为250-500μm)发生颅神经,皮肤和粘膜缺血性并发症的情况。与同尺寸的聚乙烯醇颗粒相比,鉴于微球颗粒的可压缩性和更高的渗透潜力,我们建议将水凝胶微球颗粒的尺寸过大,以进行头颈部栓塞。