Comparison of a new minimum contact locking plate and the limited contact dynamic compression plate in an osteoporotic fracture model

机译:骨质疏松性骨折模型中新型最小接触锁定板和有限接触动态压缩板的比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The purpose of this study was to investigate the biomechanical properties of a new minimum contact locking plate (MC-LP) compared with the limited contact dynamic compression plate (LC-DCP). Eighteen pairs of fresh human osteoporotic cadaver radii were equally divided into three groups. Each specimen was tested in each of the following force applications: anteroposterior (AP) four point bending, mediolateral (ML) four point bending, and torsion. A 10-mm gap osteotomy model was used to simulate a comminuted diaphyseal radial fracture. For each pair, one radius received a limited contact dynamic compression plate (LC-DCP) and the contralateral radius was fixed with a minimum contact locking plate (MC-LP). Specimens were tested in nondestructive four point bending and torsion on an electronic universal material testing system. The results indicate that the MC-LP system is significantly more stable than the LC-DCP system when tested in four point bending and torsion in an osteoporotic comminuted radial diaphyseal fracture model.
机译:这项研究的目的是调查与有限接触动态压缩板(LC-DCP)相比,新型最小接触锁定板(MC-LP)的生物力学性能。将十八对新鲜的人骨质疏松尸体半径平均分为三组。在以下每种力作用下测试每个样本:前后(AP)四点弯曲,内侧(ML)四点弯曲和扭转。使用10毫米间隙截骨模型来模拟干骨干phy骨骨折。对于每一对,一个半径接受一个有限的接触动态压缩板(LC-DCP),对侧半径用一个最小接触锁定板(MC-LP)固定。样品在电子通用材料测试系统上进行了无损四点弯曲和扭转测试。结果表明,在骨质疏松性粉碎性骨干骨折模型的四点弯曲和扭转试验中,MC-LP系统比LC-DCP系统更稳定。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号