Although considered the ‘gold standard’ for measuring blood pressure in laboratory animals, telemetry would benefit from refinement. In the present study, we tested the hypothesis that the small telemetric device used for blood pressure recording in mice would work for rats as well and would serve as an alternative for those studies where abdominal cavity space is quite limited (such as in young animals and pregnant females). Here we report that the use of a smaller and lighter telemetric device implanted in the abdominal aorta of rats led to acquisition of stable and high-quality blood pressure and heart rate data, similar to those obtained by using a larger telemetric device developed for rats. The use of smaller transmitters represents an alternative telemetry technique, especially for those cases in which space in the abdominal cavity is particularly limited such as during pregnancy.
机译:尽管被认为是测量实验动物血压的“黄金标准”,但遥测技术将受益于完善。在本研究中,我们检验了以下假设:用于小鼠血压记录的小型遥测设备也适用于大鼠,并且可以替代那些腹腔空间非常有限的研究(例如,在幼小动物和怀孕的女性)。在这里,我们报告说,使用较小且较轻的遥测设备植入大鼠腹主动脉导致获得稳定且高质量的血压和心率数据,类似于使用针对大鼠开发的较大遥测设备获得的数据。使用较小的发射器代表了另一种遥测技术,特别是对于那些在怀孕期间腹腔空间特别有限的情况。