Intra-examiner reliability of measurements of ankle range of motion using a modified inclinometer: a pilot study

机译:使用改进的测斜仪测量踝关节运动范围的检查者内部可靠性:一项初步研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A modified inclinometer was designed for measuring total ankle range of motion (ROM) in the standing position for a large future study. The purpose of this pilot study was to assess the intra-examiner reliability of this new device in order to see if the examiner would be able to produce equally reliable measurements with this instrument as with a routinely used goniometer.Nineteen young healthy individuals took part in the pilot. The same examiner took the ROM measurements using both devices twice on the same day and one further time 2 or 3 days later. Test-retest reliability was measured using the intraclass correlation coefficient (ICC). The ICC values were 0.86 (95% CI=[0.67; 0.94]) and 0.83 (95% CI=[0.61; 0.93]) for the measurements taken with the goniometer on the same day and for those on two different days. The corresponding values for the modified inclinometer were 0.88 (95% CI=[0.72;0.95]) and 0.81 (95% CI=[0.57; 0.92]). Both instruments were found to have very good test-retest reliability.
机译:设计了一种改进的测斜仪,用于测量站立位置的总踝关节运动范围(ROM),以备将来研究之用。这项初步研究的目的是评估该新设备的检查者内部可靠性,以查看检查者是否能够使用此仪器产生与常规使用的测角仪同样可靠的测量结果。十九名年轻健康的人参加了飞行员。同一位检查员在同一天使用这两种设备进行了ROM测量,并在2或3天后又进行了一次测量。使用组内相关系数(ICC)来测量重测信度。对于同一天和两天中使用测角计进行的测量,ICC值分别为0.86(95%CI = [0.67; 0.94])和0.83(95%CI = [0.61; 0.93])。改进的测斜仪的相应值为0.88(95%CI = [0.72; 0.95])和0.81(95%CI = [0.57; 0.92])。发现这两种仪器都具有很好的重测可靠性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号