Volume Measurement of Intracranial Aneurysms from 3D Rotational Angiography: Improvement of Accuracy by Gradient Edge Detection

机译:3D旋转血管造影颅内动脉瘤的体积测量:梯度边缘检测提高精度

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Summary: Manual volume measurement of intracranial aneurysms from 3D rotational angiography varies on different threshold settings and, therefore, is operator-dependent. We developed and validated a method based on automatic gradient edge detection that is independent on threshold settings and provides an accurate and reproducible volume measurement. This method was compared with manual volume calculation in 13 aneurysm phantoms, and the results were significantly more accurate.
机译:概述:3D旋转血管造影颅内动脉瘤的手动体积测量在不同的阈值设置上变化,因此是操作员所依赖的。我们开发并验证了一种基于自动梯度边缘检测的方法,该方法独立于阈值设置,并提供准确和可重复的体积测量。将该方法与13个动脉瘤素的手动计算进行了比较,结果明显更准确。