Procedure for optimal implementation of automatic tube potential selection in pediatric CT to reduce radiation dose and improve workflow

机译:儿科CT中自动管势选择的最佳实施过程减少辐射剂量提高工作流程

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

It is important to employ radiation dose reduction techniques in pediatric computed tomography (CT) to reduce potential risks of radiation‐induced malignancy. Automatic tube potential (kV) selection tools have been developed and become available on many CT scanners, which select the optimum kV based on the patient size and clinical task to improve the radiation dose efficiency. However, its use in pediatric CT has been mostly empirical, following manufacturer’s default recommendation without solid demonstration for quality improvement. This study aimed to implement an automatic tube potential tool (CAREkV, Siemens Healthcare) into routine pediatric CT practice, using the “Plan‐Do‐Study‐Act” quality improvement process, in place of an existing kV/mAs technique chart. The design of this quality improvement project involved Plan‐Do‐Study‐Act stages. Plan and Do stages identified the criteria for optimal automatic kV selection; a range of phantoms representing typical pediatric groups were scanned on a dual‐source 128‐slice scanner using a fast‐pitch scanning mode. The identified CAREkV settings were implemented into the CT protocol and evaluated after a 6‐month period. In the Study stage, an objective evaluation of the image metrics and radiation dose for two similar patient cohorts using CAREkV and the technique‐chart, respectively, were compared. The kV selected, image quality and radiation dose determined by CAREkV were comparable to those obtained while using the technique‐chart. The CAREkV was successfully implemented into our pediatric abdominopelvic CT practice. By utilizing the “PDSA” process optimal image quality and radiation dose reduction were achieved with an automatic kV selection tool to improve CT workflow.
机译:在儿科计算机断层扫描(CT)中使用辐射剂量降低技术是重要的,以降低辐射诱导的恶性肿瘤的潜在风险。已经开发了自动管势(KV)选择工具并在许多CT扫描仪上可用,基于患者大小和临床任务选择最佳KV,以提高辐射剂量效率。然而,它在儿科CT中的使用主要是经验,遵循制造商的默认建议,没有稳固示范质量改进。本研究旨在利用“计划执行法”质量改进过程,实现自动管潜在工具(CareKV,Siemens Healthcare)进入常规儿科CT实践,代替现有的KV / MAS技术图表。这种质量改进项目的设计涉及计划 - DO-STAIL-ACT阶段。计划和阶段确定了最佳自动kV选择的标准;使用快速俯仰扫描模式在双源128切片扫描仪上扫描代表典型小儿组的一系列幽灵。已识别的CareKV设置在CT协议中实施,并在6个月内进行评估。在研究阶段,比较了使用Carekv和技术图表的两种类似患者群体的图像度量和放射剂量的客观评估。由CareKV确定的KV选择,图像质量和辐射剂量与使用技术图表的同时获得的kV。 Carekv成功地实施到了我们的儿科腹腔肝癌CT实践中。通过利用“PDSA”过程,通过自动KV选择工具实现最佳图像质量和辐射剂量,以改善CT工作流程。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号