摘要

Mechanisms of ceramic liner fractures have not yet been fully elucidated. Impingement between a stem and ceramic liner is a proposed cause of ceramic liner fractures. We experienced a case of ceramic liner fracture caused by direct impingement between the stem shoulder and the ceramic liner. This type of impingement, unlike impingements with a stem neck, has not been previously reported. While we assume that certain characteristics of the stem contributed to the impingement, we report this case to note that caution may be needed when using certain stem designs.
机译:陶瓷衬垫骨折的机制尚未完全阐明。杆和陶瓷衬里之间的冲击是陶瓷衬垫骨折的提出原因。我们经历了由杆肩部和陶瓷衬垫之间的直接冲击引起的陶瓷衬垫骨折的情况。与用茎颈的切割不同,这种类型的冲击尚未报道。虽然我们假设茎的某些特性有助于调查,但我们报告了这种情况,以注意使用某些茎设计时可能需要谨慎。